Safety Notes

Using the Laparoscopy Indication
43
en
8 Using the Laparoscopy Indication
Intended use: LaparoscopyThe devices PG130/PG145 may serve as irrigation pump for use in minimally in-
vasive laparoscopic surgery (keyhole surgery). The optional suction function can
be used to siphon off secretions. The optional suction function can be used to ex-
tract and remove waste gas when using laser or RF surgical devices.
ContraindicationsThe devices may not be used in the laparoscopy indication for surgical proce-
dures where a specific pressure must be preset and reached as required such as
during hysteroscopy, urology, and arthroscopy procedures, for example.
Technical application scope: LaparoscopyThe following characteristics apply to the laparoscopy indication:
The devices PG130/PG145 work with a permanently preset pressure of
500 mmHg when used in the laparoscopy indication.
8.1 Device-Inherent Dangers: Laparoscopy Indication
WARNING!
Connecting the tube
The tube outlet may only be connected to instruments which are intended for in-
tra-abdominal fluid suction and irrigation.
8.2 Switching Device On
Switch device on1. Make sure no tube set is inserted into the tube retainer before switching the
device on.
2. Press the ON switch. The device turns on. The touchscreen displays the fol-
lowing in succession: Company logo/Device check and Company logo/Device
OK.
3. After the successful device check, the available indications are displayed for
selection (see Chapter 6.3 Activating a New Indication). If the device check is
unsuccessful, please read Chapter 6.1.1 Device Displays after Unsuccessful
Completion of Device Check).
4. Press the [--] function field. The indication screen Laparoscopy is displayed.
8.3 Screen Displays of the Laparoscopy Indication
The [STANDARD] flow level (factory setting) is set when activating the laparosco-
py indication. The [STANDARD] flow level can be set in the user menu to a value
LAP
ARTHRO
HYS URO
4.
Fig. 8-1 Laparoscopy indication screen
(1) Indication
(2) [Standard] function field
(3) [High Flow] function field
(4) Suction function indicator (only
PG145)
(5) [Indication selection] symbol
(6) [User menu] symbol
(7) [Fluid bag] symbol (optional)
(8) Status line for messages
(9) [START]/[STOP] function field
LAP
Insert tube set
START
Standard
High Flow
0
2
(1)
(5)(6)(7)(8)
(4)
(2) (3)
(9)
OBJ_DOKU-0000006143-008.fm Seite 43 Mittwoch, 11. November 2020 2:29 14