Safety Notes

42
Using Suction Function (Only Pg145)
en
7.3 Stopping Suction
Stop suction
Close the inflow valve of the instrument. On the indication screen, press the [0]
symbol to deactivate suction level 1 or 2.
7.4 Replacing the Secretion Container
CAUTION!
Full secretion containers must be replaced immediately without stopping sur-
gery. If the overflow protection of the secretion containers is triggered, suction
is stopped to prevent the ingress of fluids.
Container change during surgery The full secretion container can be replaced with an empty one during surgery.
1. Stop the suction (see Chapter 7.3 "Stopping Suction").
2. Replace the full secretion container with an empty one.
3. Start the suction (see Chapter 7.2 "Starting Suction").
7.5 Removing a Suction Tube
1. Remove the suction instrument tube (3) from the instrument (4).
2. Remove the secretion container (2) with the suction instrument tube (3).
3. Remove the secretion container tube (1) with the secretion container (2).
4. Remove the secretion container tube (1) from the device.
7.6 Safety Functions
Defective vacuum pump
If the suction function or vacuum pump is malfunctioning, the status line depicts
the following warning message in red: Vacuum pump defective! A long warning
beep is emitted 3 times. Surgery can be continued. After surgery, safeguard the
device until checked by an authorized service technician (see Chapter 15.4 "Aes-
culap Technical Service").
ARTHRO
SET
PRESSURE
45 mmHg
SET
FLOW
1.5 l/min
ACTUAL PRESSURE
0 50 100 150 200
START
Start pump
Knee
0
2
ARTHRO
SET
PRESSURE
45 mmHg
SET
FLOW
1.5 l/min
ACTUAL PRESSURE
0 50 100 150 200
START
Vacuum pump defective!
Knee
0
2