Safety Notes
Operating the Device
39
en
6.15 Using a Foot Pedal (Optional)
CAUTION!
The foot pedal can be disinfected by wiping off and must not be sterilized.
Using the foot pedalIn the arthroscopy indication, the physician can operate the device with an op-
tionally connectable foot pedal (see Chapter 20 Article/Accessory List PG130/
PG145) and thus activate and deactivate the Wash function. The device can be
controlled with the following control elements on the foot pedal:
Explanation: [x] = Function available [-] = Function not available
Starting and stopping the WASH function
via foot pedal
1. The foot pedal must be connected and the device must have recognized the
foot pedal after being switched on.
2. Press the foot pedal to start the Wash function. Please note that the Wash
function stops automatically after 10, 20 or 30 seconds and the device re-
turns to the previously specified nominal values. The remaining time of the
WASH function is depicted on the upper right of the screen in the form of a
countdown. After stopping the Wash function, it can be restarted again.
3. Press the foot pedal to stop the WASH function before the 10, 20, or 30 sec-
onds expire. After stopping the Wash function, it can be restarted again.
Fig. 6-10 Foot pedal elements
(1) Connection cable
(2) Foot pedal
(2)(1)
Buttons Function LAPAROSCOPY ARTHROS-
COPY
HYSTEROS-
COPY
UROLOGY
[WASH]
Pedal
Starting and stopping
the Wash function
-x--
OBJ_DOKU-0000006143-008.fm Seite 39 Mittwoch, 11. November 2020 2:29 14