Safety Notes
34
Operating the Device
en
6.8.1 Replacing the Patient Tube After Surgery
WARNING!
The day tube remains connected to the device after each surgery. The patient
tube must be disposed of immediately after surgery. A new sterile cap (included
with the patient tube) must be screwed onto the hygiene connector at the day
tube after each surgery. This sterile protector remains connected to the day tube
until the next surgery.
1. To be carried out by non-sterile technician:
Open packaging of the patient tube.
A sterile technician then removes the patient tube from the inside of the
package.
2. To be carried out by sterile technician:
Keep the luer lock connector (15) and hand the other end of the patient tube
to the non-sterile technician.
Connect the luer lock connector (15) of the patient tube with the instrument
(e.g., inflow cannula).
3. To be carried out by non-sterile technician:
Remove the sterile protector of the day tube and connect the new patient
tube to the day set without delay.
6.9 Removing a Tube Set
Remove tube set In order to separate sterile from non-sterile areas, assign the following tasks to
the "sterile" or "non-sterile" nurse or medical technician. The indicated numbers
refer to Fig. 6-2 Tube set elements and Fig. 6-3 Inserting the tube set.
1. To be carried out by sterile technician:
Disconnect the Luer lock connector (10) of the instrument (e.g. inflow can-
nula).
Hand the Luer lock connector (10) to the non-sterile technician.
2. To be carried out by non-sterile technician:
Disconnect the end with the tap spikes (9) from the irrigation fluid bags
(12).
Press the mechanical tube release (17) and keep depressed.
Remove the tube holder (2) with the pump segment (1) from the tube re-
tainer (14).
6.10 Attaching and Connecting the Fluid Bags
WARNING!
Fluid bags
The device is only intended for use with flexible fluid containers. Do not use glass
containers as they might break. Fluid cannot flow quickly enough due to the vac-
uum being generated inside of the bottle. Risk of implosion.
WARNING!
The physician must determine an irrigation fluid suitable for the application and
medical procedure!