Connect the luer lock connector (15) with the instrument (e.g., inflow can-
nula). Open inflow valve at instrument.
Insert new day tube3. To be carried out by non-sterile technician:
Connect the hygiene connector (13) with the patient connector (10).
Switch the device on and select the arthroscopy indication (see Chapter 9.4
"Switching Device on When Using Day Patient Set").
Select the desired joint type. The indication screen shows the following: In-
sert tube set.
Place the pump segment (1) loosely over the roller wheel (21) and press the
tube holder (2) towards the rear and into the tube retainer (19) of the de-
vice.
Press on tube holder (2) center using your thumb and push holder down ver-
tically until it snaps into the tube retainer. This can be heard and felt. The
pump segment (1) is now clamped in place firmly over the roller wheel (21).
The validity of the tube set is now checked. In case of an invalid tube set, the
status line alternately shows the following: Tube set not valid and Please re-
place. A short acoustic alarm is emitted 3x. The entire procedure must be re-
peated with a valid tube set if an invalid tube set has been inserted.
Connect the tube end with the tap spikes (8)/Care lock (7) and the full irri-
gation fluid bags.
If the Fluid bag warning function is in the General settings -> Fluid bags
menu is set to ON, press the [Fluid bag] symbol to reset the counter to 0.
Press the [START] function field and wait until the day patient set is com-
pletely filled with irrigation fluid.
The device is now ready for use.
WARNING!
The physician must determine an irrigation fluid suitable for the application and
medical procedure!
Inserted day tube3. To be carried out by non-sterile technician:
Connect the hygiene connector (13) with the patient connector (10).
Switch the device on and select the arthroscopy indication (see Chapter 9.4
"Switching Device on When Using Day Patient Set").
Select the desired joint type. The indication screen shows the following: Day
tube inserted.
The validity of the tube set is now checked. In case of an invalid tube set, the
status line alternately shows the following: Tube set not valid and Please re-
place. A short acoustic alarm is emitted 3x. The entire procedure must be re-
peated with a valid tube set if an invalid tube set has been inserted.
Connect the tube end with the tap spikes (8)/Care lock (7) and the full irri-
gation fluid bags.
Press the [Fluid bag] symbol to reset the counter to 0.
Press the [START] function field and wait until the day patient set is com-
pletely filled with irrigation fluid.
The device is now ready for use.
WARNING!
The physician must determine an irrigation fluid suitable for the application and
medical procedure!