Safety Notes

Operating the Device
31
en
6.8 Inserting the Day Patient Tube Set
The day patient set consists of two tube sections comprised of the day tube and
the patient tube. The day tube establishes the tube connection between the irri-
gation fluid bags and the patient tube. The day tube is inserted into the tube re-
tainer of the device and not replaced for the duration of one day between
surgeries and/or surgical procedures. Only the patient tube (tube connection be-
tween day tube and the instrument) must be replaced between surgical proce-
dures. Efficiency is increased with a day set if several surgeries are to be
performed on one day. The patient tube has to be replaced after each surgical
procedure. Make sure to connect tubes as indicated by the matching connectors.
CAUTION!
Use only original day patient sets by the manufacturer. Other day patient set sys-
tems are not compatible with the pump!
The day tubeThe day tube includes the connectors to the fluid bags and to the hygiene con-
nector (Fig. 6-5 "Patient tube elements" see (15)) of the patient tube.
The day tube may be used the entire day of surgery. However, it is designed for a
maximum of 10 uses. After 10 uses, the day tube must be replaced with a new
one.
WARNING!
The day tube must be replaced at the end of the day of surgery but not later than
after 24 hrs.
A new sterile cap (included with the patient tube in the delivery) must be screwed
onto the patient connector (Fig. 6-4 (10)) of the day set immediately after the
surgical procedure to prevent contamination. This sterile protector remains con-
nected to the day tube until the next surgery.
The patient tubeThe patient tube has a connector on one side to connect with the day tube (hy-
giene connector (13)). A luer lock connector (15) to connect to an instrument
Fig. 6-4 Day tube set elements
(1) Pump segment
(2) Tube holder with pressure mem-
brane
(3) Transponder
(4) Inflow tube
(5) Y-connector
(6) Tube clamps
(7) Care lock connectors
(8) Tap spikes
(9) Protective caps
(10) Patient connection (Care lock)
(11) Instrument tube, 15 cm
(2)
(5)
(6)
(7)
(1)
(3)
(4)
(8)
(9)
(10)
(11)