Safety Notes

30
Operating the Device
en
In order to separate sterile from non-sterile areas, assign the following tasks to
the "sterile" or "non-sterile" nurse or medical technician. The indicated numbers
refer to Fig. 6-2 "Tube set elements" and Fig. 6-3 "Inserting the tube set".
Open outer packaging 1a. Disposable tube set - To be carried out by non-sterile technician:
Open outer packaging of the tube set.
A sterile technician then removes the inner tube set package and opens it.
Open sterile autoclavable container 1b. Reusable tube set - To be carried out by non-sterile technician:
Open the sterile autoclavable container of the tube set.
A sterile technician then removes the tube set from the inside of the pack-
age.
Connect instrument 2. To be carried out by sterile technician:
Keep the Luer lock connector (10) in the sterile area and hand the tube end
with the tap spikes (9)/Care lock (7) to the non-sterile technician.
Connect the Luer lock connector (10) with the instrument (e.g. inflow can-
nula). Open inflow valve at instrument.
Insert tube set 3. To be carried out by non-sterile technician:
Turn the device on and select the desired indication. The indication screen
shows the following: Insert tube set.
A. Place the pump segment (1) loosely over the roller wheel (16) and press
the tube holder (2) towards the rear and into the tube retainer (14) of the
device.
B. Press on tube holder (2) center using your thumb and push holder down
vertically until it snaps into the tube retainer. This can be heard and felt. The
pump segment (1) is now tensioned firmly over the roller wheel (16).
The validity of the tube set is now checked. In case of an invalid tube set, the
status line alternately shows the following: Tube set not valid and Please re-
place. An acoustic alarm is emitted 3x. The entire procedure must be repeat-
ed with a valid tube set if an invalid tube set has been inserted.
Connect the tube end with the tap spikes (9)/Care lock (7) and the full irri-
gation fluid bags (12).
If the Fluid bag warning function is enabled (see "Setting the Fluid Bag Warn-
ing"), press the [Fluid bag] symbol to reset the device to a full fluid bag. Make
sure the bag volume is set correctly.
Press the [START] function field and wait until the tube set is completely
filled with irrigation fluid.
The device is now ready for use.
Fig. 6-3 Inserting the tube set
(12) Irrigation fluid bags
(13) Tap spikes/Care lock
(14) Tube retainer
(15) Inflow tube/irrigation tube
(16) Roller wheel
(17) Mechanical tube release
(18) Instrument tube
!
(13)
(14)
(15)
(18)
(12)
(16)
(17)
A.
B.