Safety Notes
28
Operating the Device
en
device is switched off or in case of a power failure, the current use cycle is in-
validated. Once the last use cycle has been invalidated, irrigation can no longer
be started. If this is the case, you must insert a new, valid, and allowed day
tube.
Transponder signal loss The device stops and cannot be restarted if the transponder loses its signal dur-
ing current operation (e.g. defective electronic component). If the signal is re-
stored with 60 seconds, it is possible to continue using the tube set.
6.6 Usable Tube Sets Overview
NOTE!
Tubes must be stored at room temperature. The shelf life for all tubes is 5 years.
NOTE!
Observe applicable hygiene regulations when disposing of the tube set.
A total of 9 different tube sets can be inserted into the tube retainer on the front
of the device. The following table lists the function of each type of tube set de-
pending on the selected indication. Explanation: [x] = usable, [-] = not usable.
WARNING!
Disposable tube sets contain Diethylhexylphthalate (DEHP), which is classified as
toxic to reproduction according to the EU Directive 67/548/EEC on Classification
and Labelling of Dangerous Substances. DEHP may impair fertility, may cause
harm to unborn child, may excrete in breast milk. Therefore, this product must
not be used for unauthorized applications. When applied within the intended
use, the potential risk to pregnant or breastfeeding women as well as to children
resulting from the DEHP contained in this product is not critical. In regard to the
short exposure time and the physical characteristics the eventuality of critical
quantities of DEHP being dissolved from the tube sets is neglectable.
Tube type Art. No. Tube category LAP ARTHRO HYS URO
Reusable PG131 Reusable tube set, luer lock con-
nector, and tap spikes
xx x x
Disposable (one-time use) PG132SU Disposable tube set, luer lock con-
nector, and tap spikes
xx x x
PG133SU Disposable tube set, luer lock con-
nector, and Care lock
xx x x
LAP tube set PG122SU Disposable tube set with suction/
irrigation grip, luer lock connector,
and tap spikes
x- - -
PG123SU Disposable tube set with suction/
irrigation grip, luer lock connector
and Care lock
x- - -
Day patient set PG134SU Day patient set with luer lock con-
nector and tap spikes, consisting
of day tube and patient tube
-x - -
PG135SU Day patient set with luer lock con-
nector and Care lock, consisting of
day tube and patient tube
-x - -
Day tube PG136 Day tube with luer lock connector
and tap spikes
-x - -
PG137 Day tube with luer lock connector
and Care lock
-x - -
OBJ_DOKU-0000006143-008.fm Seite 28 Mittwoch, 11. November 2020 2:29 14