Safety Notes
26
Operating the Device
en
1. Switch device on. The touchscreen of the device depicts the language selec-
tion if an indication has not yet been activated. Press the desired language.
2. Press [Save] to save the language.
3. The touchscreen depicts the following in the selected language: [Indication
activation]. Press [Indication activation]. Please follow the instructions on
the touchscreen.
4. Hold the desired transponder onto the center of the horizontal area under-
neath the roller wheel (see Fig. 6-1 "Inserting transponder into tube retain-
er").
5. The device checks the reliability and validity of the transponder. With a suc-
cessful check, the indication screen depicts the activated indication. All other
indications can now be activated as described in Chapter 6.3 "Activating a
New Indication".
If the transponder is invalid, the screen depicts the following: Activation tran-
sponder invalid. If this is the case, repeat the process with a new and valid
transponder.
6.3 Activating a New Indication
Activate new indication with transponder A device with already one or several activated indications, can be enabled for one
additional indication (max. of 4 indications per device!). Activation is done indi-
vidually for each indication using a special transponder (see Chapter 20 "Article/
Accessory List PG130/PG145").
Invalidating a transponder Once an indication has been activated with a transponder, that transponder is
automatically invalidated and can not be used a second time. Each transponder
can be used only once.
Activate new indication Up to 4 indications can be activated for the devices PG130/PG145.
1. Switch device on.
2. All of the already activated indications are depicted on the touchscreen. Press
the function field of one of the depicted indications (e.g., [ARTHRO]). The in-
dication screen opens. If you have selected the indication ARTHRO, then you
must press one of the displayed function fields (e.g., [Knee]) before being able
to press the [User menu] symbol.
3. Press the [User menu] symbol. The corresponding user menu opens.
4. Press [General settings]. The following function field opens, among others:
[Indication activation].
5. Press the [Indication activation] function field. Please follow the instructions
on the touchscreen. The device checks the reliability and validity of the tran-
sponder. With a successful check, the indication screen depicts the new acti-
vated indication and all previously activated indications.
If the transponder is invalid, the screen depicts the following: Transponder inval-
id. If this is the case, repeat the process with a new and valid transponder.
Fig. 6-1 Inserting transponder into
tube retainer
(1) Tube retainer
(2) Transponder
!
AR
THRO
(1)
(2)
Deutsch
Language
English Fran aisç
Español Italiano
1.
2.
Indication
activation
3.
ARTHRO
SET
PRESSURE
45 mmHg
SET
FLOW
1.5 l/min
ACTUAL PRESSURE
0 50 100 150 200
START
Insert tube set
Knee
3.
Fluid bags
Volume
Indication activation
Menu
General settings
Brightness
5.