Safety Notes

Operating the Device
25
en
6 Operating the Device
6.1 Turning the Device On
1. Connect the device with the mains power.
2. Make sure no tube set is inserted into the tube retainer (exception: day set
during arthroscopy! See 9.4 "Switching Device on When Using Day Patient
Set", item b). Press the ON switch. An acoustic warning signal is emitted.
3. The device performs a device check after being switched on. The touchscreen
first displays the Company logo/Device check. After approx. 2 seconds, Com-
pany logo/Device OK is displayed on screen. A beep is emitted 3 times. After
the successful device check, the screen depicts all activated indications in the
form of function fields. A maximum of 4 indication types are available:
LAP for laparoscopy
ARTHRO for arthroscopy
HYS for hysteroscopy
URO for urology
4. If a tube set is in the tube retainer before the device is being switched on, the
screen depicts the following message: Remove tube set (exception: day set
during arthroscopy! See 9.4 "Switching Device on When Using Day Patient
Set", item b). Remove tube set. The device conducts a device check (see item
3).
If the device check is unsuccessful, the corresponding error messages are dis-
played (see Chapter 6.1.1 "Device Displays after Unsuccessful Completion of
Device Check").
6.1.1 Device Displays after Unsuccessful Completion of Device Check
If the device check was not successful, the touchscreen depicts the respective de-
vice error. The following errors may be depicted:
Please read Chapter 18 "Error and Warning Messages" for additional informa-
tion.
6.2 Activating Indication on New Factory Devices
Activate indication with transponderDevices new from the factory are shipped without any activated indications. The
desired indication must be activated before first use. Activation is done individu-
ally for each indication using a special transponder (see Chapter 20 "Article/Ac-
cessory List PG130/PG145").
The activated indications are depicted on the touchscreen in color. Not activated
indications are depicted as gray areas.
Invalidating a transponderOnce an indication has been activated with a transponder, that transponder is
automatically invalidated and can not be used a second time. Each transponder
can be used only once.
Device error:
Motor error
Sensor error
Electronic error
Calibration error
!
Device check
Remove tube set
LAP
HYS