Safety Notes
Technical Data
113
en
* When using the day patient set and a pressure of 300 mmHg for max. 16 sec-
onds, the day patient set is identified as contaminated.
Essential Performance Hysteroscopy: Pressure build-up in the body cavity, control and mea-
surement, limit value: max. 150 mmHg (normal state)
First error: 200 mmHg for max. 5 seconds
Urology: Pressure build-up in the body cavity, control and mea-
surement, limit value: max. 90 mmHg (normal state)
First error: 100 mmHg for max. 5 seconds
Arthroscopy: Pressure build-up in the body cavity, control and mea-
surement, limit value: max. 200 mmHg (normal state)
First error: 250 mmHg for max. 5 seconds *
Laparoscopy: Generate irrigation flow, 1.0 l/min, 1.4 l/min, ± 10 %
(normal state)
First error: 0 l/min (no function) or 3.5 l/min (limit due
to design constraints)