Safety Notes
112
Technical Data
en
Adjustable values/ranges Nominal flow range: Arthroscopy: 0.1-2.5 l/min (max. flow)
Hysteroscopy: 50-500 ml/min (max. flow)
Urology: 25-500 ml/min (max. flow)
Laparoscopy: 1.0-3.5 l/min (max. flow)
Suction power: Max. 2.0 l/min (depending on outflow cannula)
Starting nominal pressure range: Arthroscopy: 15-200 mmHg
Hysteroscopy: 15-150 mmHg
Urology: 15-90 mmHg
Laparoscopy: not adjustable
Deficit threshold: Hysteroscopy: 100-3000 ml
Measurement range Flow: 0-3.5 l/min
Pressure: 0 to 750 mmHg
Deficit/inflow volume: 9995 ml
Precision (repeatability) Flow: ± 10 %
Pressure: ± 10 %
Deficit: ±10 % (relative to pumped inflow volume)
Accuracy Flow: ± 10 %
Pressure: ± 5 % (of final value)
Deficit/inflow volume: ±10 % (relative to pumped inflow volume)
Dimensions Width x Height x Depth: 305 x 183 x 305 [mm]
12.0 x 7.2 x 12.0 [inch]
Mass: 8.4 kg
Interfaces/ports IN/OUT signal for components: 1 scale connection (RS232 socket, DSUB9/RS232)
1 service port (USB port, USB 2.0)
1 foot pedal (acc. to IEC 60601-1)
Mains power socket: IEC 60320-1 C14
RFID Transponder Technology
Transmit/Receive Frequency Range: 13.56 MHz
± 0.424 MHz
Transceiver class: Class I
RF Output Power: -10.83 dBμA/m at 10 m/ 32.8 ft
Type of Antenna: Inductive Loop Antenna
Antenna loop area: 0.00032 m²
Modulation: Amplitude-shift keying (ASK)
Mode of Operation (Simplex / Duplex): Duplex
for use with remote control PG125 - Infrared interface