Safety Notes

Technical Data
111
en
19 Technical Data
Type designation: PG130/PG145
Production date: AESCULAP AG
Am Aesculap-Platz
D-78532 Tuttlingen
Germany
Software version: See instructions for use to determine software version (Chapter 12.3.1 "Software Version").
Mains voltage range: 100-240 V~
Supply frequency range: 50/60 Hz
Fuse designation: 2 x T 3,15 AH, 250 V, UL-recognized, slow blow
Internal voltage supply: No
Max. power consumption: 125 VA
Max. current consumption: 100 V: 1150 mA
240 V: 510 mA
Protection class (I, II, III) I
Application part type [B, BF,
CF]:
BF
Defibrillator protected appli-
cation part:
No
Type of protection (IP code): IP41
Classification according to
Medical Device Directive 93/
42/EEC, (I, IIa, IIb:)
IIb
Tested to the following stan-
dards (in the respectively
valid version):
EN 60601-1/IEC 60601-1
EN 60601-1-2/IEC 60601-1-2
Operating conditions: 10-40° C / 50-104° F
30 to 75% rel. humidity
70-106 kPa air pressure
3000 m max. above sea level for device use
Possible use with explosive
anesthetic gases
This system is not designed for use with flammable anesthetic gases (Class AP) or flammable
anesthetic gases with oxygen (Class APG).
Storage conditions: 5-40° C / 41-104° F
5 to 85 % rel. humidity
70-106 kPa air pressure
Transport conditions: -20 to +60° C / -4 to 140 °F
10-90 % rel. humidity at 30 °C/86 °F
70-106 kPa air pressure
Max. sound level: 65 dB(A)
80 dB(A) (with acoustic signals)
Max. scale capacity: 7 kg
Maximum output pressure: 500 mmHg ± 10 %
Negative suction pressure,
vacuum pump:
Max. 700 mbar
OBJ_DOKU-0000006143-008.fm Seite 111 Mittwoch, 11. November 2020 2:29 14