Initial Device Startup
en
4 Initial Device Startup
Operation of the devices PG130/PG145 is reserved for persons with the relevant
professional qualifications and trained to use the device.
4.1 Delivery Scope
Device PG130/PG145
Instruction for use
Power cable
Delivery inspection Always check all parts and optional accessories of the device immediately after
receiving the shipment. The manufacturer considers only replacement claims
that have been immediately submitted or reported to a sales representative or an
authorized service company.
Returning the device Use the original packaging material if the device has to be returned. The manu-
facturer is not responsible for transport damages due to insufficient or unsuit-
able packaging.
Please make sure that all required information has been supplied:
•Name of owner
•Address of owner
Device type and model
Serial number of the equipment (see identification plate)
Detailed description of defect
4.2 Setting up and Connecting the Device
NOTE!
Locations
The device may be used only in the hospital or operating rooms.
Setting up Place the device on a flat surface free of vibration located in a dry environment.
The ambient temperature and humidity must meet the requirements mentioned
in Chapter 19 "Technical Data".
WARNING!
ME System (Medical Electrical System)
Use only parts and/or devices from ME systems (see Chapter 22 Glossary) in pa-
tient environments in compliance with the standard IEC60601-1 in the respec-
tive currently valid version.
WARNING!
Ventilation of the device
Avoid device overheating. Ensure free air circulation especially to the bottom
and rear of the device (rear panel distance of at least 10 cm).
CAUTION!
ME systems (Medical Electrical System)
The medical electrical device is suitable for integration in ME equipment systems
(see Chapter 22 Glossary). Operation of the ME device in vicinity of non-ME de-
vices may result in voiding the intended use of the ME device.