3.4.3 Density measurement
The attainable accuracy depends on the selected calibration type.
Without calibration no density measurement is possible and the empty pipe recognition
is not available!
Model
without 3-Point 5-Point
CMU-C 5 g/l 2 g/l
CMU-D 5 g/l 2 g/l
CMU-E 5 g/l 1 g/l
CMU-G 5 g/l 1 g/l
CMU-H 5 g/l 1 g/l
CMU-J 5 g/l 2 g/l
CMU-K 5 g/l 2 g/l
CMU-L 5 g/l 2 g/l
CMU-N 5 g/l 2 g/l
CMU-P 5 g/l 2 g/l
CMU-Q 5 g/l 2 g/l
CMU-R 5 g/l 2 g/l
no measurement of density
Density accuracy
3.4.4 Accuracy
Mass flow
Accuracy CMU-C to CMU-H ± 0.1% of actual flow + zero point stability (see Section 3.4.2 CMU
flow ranges)
Accuracy CMU-J to CMU-R ± 0.15% of actual flow + zero point stability (see Section 3.4.2 CMU
flow ranges)
Repeatability error ± 0.05% of actual flow (sensor with transmitter)
Additional measured values
Volume flow ± 0.2 % of actual value + zero point stability
Temperature ± 0.5 °C
Hysteresis n/a
Settling time 1 to 15 seconds
Startup drift 15 minutes
Long-term drift ± 0.02 % of upper-range value per year
Influence of ambient temperature ± 0.005 % per K
Influence of fluid temperature Compensated
Influence of fluid pressure For fluids: too small to be relevant