Setup and Menus User Manual
Commissioning
Fluid Components Intl
Document 06EN003327 Rev. -
55 FlexCOR
™ Model CMF Series
Basic settings
Menu description
Setting of min./max.
values and units
Example; programming
of max. mass flow
Keypad operation Implementation Display on Transmitter
Push for 2 sec. To access the Password
user password 0000
Push once To unlock password CHANGE
0000
Push once To enter 1000 as CHANGE
password 1000
Push once To lock password and to CONV.SETUP MODE>
enter the menu Basic settings
Push once To enter basic setting Flow direction
submenu Positive
Push once To go to mass flow Massflow max.
max. setting 000020. kg/h
Push once To change num. Massflow max.
value 000020. kg/h
Push 4 times To move the cursor Massflow max.
to the num. position 000020. kg/h
Push Until 4 appears Massflow max.
000040. kg/h
Push once To move the cursor to Massflow max.
the next num. position 000040. kg/h
Push Till 5 appears Massflow max.
000045. kg/h
Push once To move the cursor to Massflow max.
the decimal point 000045. kg/h
Push To position the decimal Massflow max.
point correct 0000.45 kg/h
Push 3 times To move cursor Massflow max.
to "Kg" unit 0000.45 kg/h
Push twice To change units to lb. Massflow max.
0000.45Lb/h
Push once To move cursor to Massflow max.
the "h" unit 0000.45Lb/h
Push 3 times To change "h" to "min" Massflow max.
0000.45 Lb/min
Push To lock the new setting Massflow max.
of the mass flowmeter 000.45 Lb/m
Push twice Transmitter reverts to
standard operation
The basic setting menu is used for basic configuration of the mass flow meter with a choice of units,
minimum and maximum limits for display and analog/digital outputs for all measurement
parameters, i.e. mass flow, volume flow, fraction, temperature and density.
Numerical values are entered by placing the cursor in the field that is to be set using the SELECT
key. Press unlock and the value can be changed using the change key. The desired value is locked
by activating LOCK.
Positioning of the decimal point is carried out by placing the cursor below the decimal point using
the SELECT key. The position can be set using the set key. The LOCK key is activated and the
decimal point is now positioned.
Selecting the unit: Place the cursor below the unit using SELECT key. Set the desired unit using
CHANGE key. Activate the LOCK key to save the setting. Place the cursor below the time scale
using SELECT key and choose the desired time scale using CHANGE, then save the desired value
by activating the LOCK key.
The maximum and minimum values set will then apply to all current and frequency/pulse outputs,
e.g. where the min. value will correspond to 0/4 mA depending on the setting of the current output
and the max. will correspond to 20 mA.
As example we want to change the default setting of the maximum mass flow on a CMF-A from
20 Kg/h to .45 lb/min.