Setup and Menus User Manual
Commissioning
FlexCOR™Model CMF Series 70
Fluid Components Intl
Document 06EN003327 Rev. -
Parameter Factory setting Settings available
Password
Password 1000 1000 – 9999
Basic settings
Flow direction Positive Positive, negative
Massflow max. Dim. dependent Dim. dependent
- Weight units Dim. dependent mg, g, kg, ton, lb
- Time units Dim. dependent s, min, h, d
Volumeflow max. Dim. dependent Dim. dependent
- Volume units Dim. dependent m
3
, ml, l, hl, kl, Ml, ft
3
, in
3
, US G, US MG,
UK G, UK MG
- Time units Dim. dependent s, min, h, d
Density
- Minimum +0.1 g/cm
3
-20000.0 kg/m
3
- +20000.0 kg/m
3
- Maximum +2.0 g/cm
3
-20000.0 kg/m
3
- +20000.0 kg/m
3
- Weight units g mg, g, kg, ton, lb
- Time units cm
3
cm
3
, m
3
, ft
3
, in
3
Sensor temp.
- Minimum -50 °C -250 °C - +250 °C
- Maximum +250 °C -250 °C - +250 °C
- Temperature units °C °C, °F, K
Fraction A/B max. Application dependent Application dependent
- Weight units kg mg, g, kg, ton, lb
- Time units h s, min, h, d
Totalizer 1 Mass flow Mass flow, fraction A, fraction B, volume flow
Forward Forward, reverse, net
Dim. dependent mg, g, kg, ton, lb
Totalizer 2 Volume flow Mass flow, fraction A, fraction B, volume flow
Forward Forward, reverse, net
Dim. dependent m
3
, ml, l, hl, kl, Ml, ft
3
, in
3
, US G, US MG,
UK G, UK MG
Low flow cut off 1.5 % 0 – 9.9 %
Empty pipe limit Dim. dependent -20000.0 g/cm
3
- +20000.0 g/cm
3
Noise filter 3 1 (min.) to 5 (max.)
Error level Warning Fatal, permanent, warning
Output
Current output 1 Off Mass flow, fraction A, fraction B, volume flow,
sensor temp., density, pct. fraction A, off
- Direction Unidirectional Unidirectional, bidirectional
- Output mode 4 – 20 mA 0 - 20 mA, 4 - 20 mA
- Time constant 5 s 0 – 30 s
Digital output 1 Pulse Pulse, frequency, error level, error number,
direction/limit, batch, off
Pulse Mass flow Mass flow, fraction A, fraction B, volume flow
- Amount/pulse Dim. dependent Dim. dependent
- Pulse polarity Positive Positive, negative
- Pulse width 66 ms 64 us, 130 us, 260us, 510 us, 1.0 ms, 2.0 ms,
4.1 ms, 8.2 ms, 16 ms, 33 ms, 66 ms, 130 ms,
260 ms, 520 ms, 1.0 s, 2.1 s, 4.2 s
Frequency Off Mass flow, fraction A, fraction B,
volume flow, sensor temp., density, pct.
fraction A
- Direction Unidirectional Unidirectional, bidirectional
- Max. frequency 10 kHz 500 Hz, 1 kHz, 5 kHz, 10 kHz
- Time constant 5 s 0 – 30 s
Error number Off 0 - 255
7.7.1 Parameters