Setup and Menus User Manual
Commissioning
Fluid Components Intl
Document 06EN003327 Rev. -
63 FlexCor
™ Model CMF Series
Example; setting the
0-point
Keypad operation Implementation Display
Push for 3 sec. To access the Password
user password 0000
Push once To unlock password CHANGE
0000
Push once To enter 1000 as CHANGE
password 1000
Push once To lock password and to CONV.SETUP MODE>
enter the menu Basic settings
Push 4 times To go to reset Reset
mode menu mode
Push once To enter the reset Totalizer 1 reset
mode menu xxx. G
Push twice To go to zero Zero adjust
adjust menu
Push once To enter the zero Zero adjust
adjust menu Auto
Push twice To go to zero Zero adjust
adjust activation +xxxxxx kg/h
Push once To activate the 0-point Zero adjust
adjustment routine +xxxxxx kg/h
Transmitter preforms Counting up from 0
0-point adjustment to 100%
New 0-point is calculated Zero adjust
and stored in SENSORPROM
â
+yyyyyy kg/h
Push twice/ Transmitter reverts to
hold 3 sec. standard operation
Menu detail