3.4 Performance characteristics of the CMB sensor
3.4.1 Reference conditions
• Established flow profile
• Inlet section has to correspond to mounting length
• Operation is to be realized in the presence of downstream control valves
• Measurement is to be realized in the absence of any gas bubbles
• Flow tubes are to be kept clean at all times
• Process temperature is to be regulated as specified in Section 3.6.1 Process temperature on
page 23
• Process pressure is to be regulated as specified in Section 3.6.6 Process pressure range on
page 23
• Ambient temperature is to range from + 10 °C to + 30 °C (50 °F to 86 °F)
• Warm-up period: 15 minutes
• Standard calibration is to be realized at 20 %, 50 % and 100 % (three times each)
• High-frequency interference is to be regulated as specified in Section 17.2 Electromagnetic
compatibility on page 104
3.4.2 CMB flow ranges
Model
kg/h [lbs/min] kg/h [lbs/min] kg/h [lbs/min] kg/h [lbs/min]
CMB-B 20 [0.7] 200 [7.3] 96 [3.5] 0.02 [0.001]
CMB-C 35 [1.3] 350 [12.9] 282 [10.4] 0.035 [0.00]
CMB-D 120 [4.4] 1,200 [44.1] 1,030 [37.8] 0.12 [0.00]
CMB-E 300 [11.0] 3,000 [110.2] 3,000 [110.2] 0.3 [0.0]
CMB-F 600 [22.0] 6,000 [220.5] 5,750 [211.3] 0.6 [0.0]
CMB-G 2,000 [73.5] 20,000 [734.9] 13,000 [477.7] 2 [0.1]
CMB-J 4,000 [147.0] 40,000 [1,469.7] 34,000 [1,249.3] 4 [0.1]
CMB-K 6,000 [220.5] 60,000 [2,204.6] 57,000 [2,094.4] 6 [0.2]
Zero point stability
(of range)
Max.
measuring range
Nominal
(p=1bar)
Min.
measuring range
Reference conditions: in conformity with IEC 770:
Temperature: 20 °C, relative humidity: 65 %, air pressure: 101.3 kPa
Fluid: water
CMB & CT OPERATING MANUAL
Part No 06EN003384 rev. 1.6
Page 15 of 106