User Manual
8
Priming Pump
CAUTION – All suction and discharge valves MUST be OPEN, as well as filter air relief valve (if available) on
filter, when starting the circulating pump system. Failure to do so could result in severe personal injury.
Release all pressure from filter, pump, and piping system. See filter owner’s manual.
If water source is higher than the pump, pump will prime itself when suction and outlet valves are opened. If water
source is lower than the pump, unscrew and remove strainer cover; fill strainer housing with water.
Clean and lubricate strainer cover O-ring with high quality O-ring lubricant each time it is removed. Inspect O-ring and
re-install on strainer cover.
Replace strainer cover on strainer housing; turn clockwise to tighten cover.
NOTE – Tighten strainer cover by hand only (no wrenches).
Turn on power and wait for pump to prime, which may take up to five (5) minutes. Priming time will depend on vertical
length of suction lift and horizontal length of suction pipe. If pump does NOT prime within five minutes, stop motor and
determine cause. Be sure all suction and discharge valves are open when pump is running. See Troubleshooting Guide.
ATTENTION – Wait five (5) seconds before re-starting pump.
Failure to do so may cause reverse rotation of motor and consequent serious pump damage.
Close filter manual air relief valve after pump is primed.
Section #5, Maintenance
Clean strainer basket regularly. Do NOT strike basket to clean. Inspect strainer cover gasket regularly and replace as
necessary. Damage caused by abuse of the strainer basket is NOT covered by your warranty.
Pumps have self-lubricating motor bearings and shaft seals. No lubrication is necessary.
Keep motor clean. Insure air vents are free from obstruction to avoid damage. Do NOT use water to hose off motor.
Occasionally, shaft seals must be replaced, due to wear or damage. Replace with genuine seal assembly kit.
See “Shaft Seal Change Instructions” in this manual.
Storage/Winterization
WARNING – Separation Hazard.
Do not purge the system with compressed air. Purging the system with compressed air can cause components to explode,
with risk of severe injury or death to anyone nearby. Use only a low pressure (below 5 PSI), high volume blower when air
purging the pump, filter, or piping.
ATTENTION – Allowing the pump to freeze will void the warranty.
ATTENTION – Use ONLY propylene glycol as antifreeze in your pool/spa system. Propylene glycol is nontoxic
and will not damage plastic system components; other anti-freezes are highly toxic and may damage plastic components in
the system.
• Drain all water from pump and piping when expecting freezing temperatures or when storing pump for a long time
(see instructions below).
• Keep motor dry and covered during storage. To avoid condensation/corrosion problems, do NOT cover or wrap
pump with plastic film or bags.