User guide

18of34 MN300282 RevA
CONFIGURATION(cont.) StepFive
HowtosetRelayFailSafe/LOST:Intheeventthesensordoesnotreceiveanecho,theFailSafeOutputor
LOST setting can be set to turn each relay ON, OFF or HOLD (to last state) to achieve the desired failsafe
condition.Duringfailsafe,thedisplaywillreadLOST.
1. InTOPLEVELMENUmode,selectSAFE.
2. SelectR1(Relay1),R2(Relay2)orR2(Relay3).
3. SelectON,OFF,orHOLD.
4. SelectEXITtoreturntoTOPLEVELMENU.
Power Outage:If power to the EchoSwitch® is
removed (intentionally or accidentally), the relays
willreturntothenormalstate.Thus,inRelay1,NC
terminalwillbeclosedandNOterminalwillbeopen.
Also,the NO terminalsin relays2 and 3will also be
open.Make sure that the deenergized state is the
safe state in your system design.The EchoSwitch®
doesusenonvolatilememorysoanypowerlosswill
notaffectthesettingsorconfigurationofthesensor.
HowtoselectFailSafeCurrentOutput/LOST:Intheeventthesensordoesnotreceiveanecho,theFailSafe
Current Output or LOST setting can be set to output a current of 4mA, 21mA or Hold (last known value).
Duringfailsafe,thedisplaywillreadLOST.
1. InTOPLEVELMENUmode,selectSAFE.
2. SelectOUTPUT.
3. Select4mA,21mA,orHOLD.
4. SelectEXITtoreturntoTOPLEVELMENU.
