Instruction Manual
14of16 MN204260 RevA
GETTINGAROUNDPODVIEW® StepEight
ChanginganAlarmSetPoint:EchoPod®sensormodelswithrelayshave(1‐4)channels.Ifafteraccessingthe
MENU, ALARMS appears in the display, then at least (1) relay is configuredas an alarm.The Alarm settings
may be in any combination of LOW and/or HIGH alarms (4‐HIGH, 1‐LOW & 3‐HIGH, 2‐LOW & 2‐HIGH, etc.).
Note: Prior to making any changes, we recommend that you write down all existing set point values.The
examplebelowhighlightsa2‐LOWand2‐HIGHalarmoperationwith(4)setpoints.Usethefollowingstepsto
changeyouralarmsettings.
StepstochangeAlarmsettings:
1. HoldSELECTfor5secondstoentertheMENU.
2. PressSELECTwhenALARMSappears.
3. PressSELECTwhenthesetpoint(HIGH2,HIGH1,LOW1,LOW2)youwanttochangeappears.
4. Press the UP and DOWN buttons to increase or decrease the set point to the desired value.To scroll
faster,holdSELECTwhilepressingUPorDOWN.
5. Toenterthesetpoint,holdSELECTfor2seconds.
6. Tochangeanothersetpoint,pressSELECTwhenthesetpointappears.
7. ToexittheMENU,pressSELECTwhenRUNappears.
Hints:
Never place a relays set point (High 1, Low 1) at the liquid empty or liquid full position.You should
haveatleastsome distanceorvolumebufferseparatingthem.Forexample,ina500gallontank,the
relaysetpointscouldbeplacedat≥10gallonsor≤
490gallons.