User Manual

8of14 LC900008 RevA
ELECTRICAL StepFive
StrobeAlertOutput(LC30‐___2Series): Withthe
StrobeAlertwiredNC;itcanbeusedasahighorlow
level alarm, depending on the setting for the invert
switch.TheStrobeAlertcanalsobewiredNO.
Changingfrom120to240VAC:
1. Remove the two screws from the top of the
printed circuit board (PCB) and gently slide the
PCB from the housing.Use caution when
removingthePCB.
2. LocatedjumpersJWA,JWBandJWConthePCB.
3. Tochangeto240VAC,removejumpersfromJWB
and JWC and place a single jumper across JWA.
To change to 120 VAC, remove jumper JWA and
placejumpersacrossJWBandJWC.
4. Gently return PCB into housing and replace the
twoscrews.
StandardVersion(NonCE)
120VAC 240VAC
Configuration Configuration
CEVersion
(LC3010_1CE/LC3010_1ECE)
120VACConfiguration 240VACConfiguration
