Specifications
Document Doc.Id. Version
Name of document GR 21-005 01
IMMUNITY LIMITS & MEASUREMENT
PHENOMENON STANDARD
TEST
METHOD
TEST LEVELS RESULT
ELECTROSTATIC
DISCHARGE
IEC 60601-1-2
IEC 61000-4-2
±2 kV, ±4 kV, ± 6 kV, ± 8 kV - Contact Discharge
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV - Air Discharge
PASS
RF ELECTROMAGNETIC
FIELDS
IEC 60601-1-2
IEC 61000-4-3
10 V/m
80 – 2700 MHz
80 % AM at 1 kHz
PASS
EL FAST TRANSIENTS
COMMON MODE / BURSTS
IEC 60601-1-2
IEC 61000-4-4
± 2 kV for A.C. Power Lines
100 kHz Burst Frequency
PASS
SURGES LINE TO LINE IEC 60601-1-2
IEC 61000-4-5
± 0.5 kV & ±1 kV At 0°, 90°, 180°, 270° ; Repetition Rate: 60 S
PASS
SURGES LINE TO GROUND IEC 60601-1-2
IEC 61000-4-5
± 0.5 kV, ±1 & ±2 kV At 0°, 90°, 180°, 270°
PASS
RF COMMON MODE IEC 60601-1-2
IEC 61000-4-6
3 & 6 V for A.C. Main Port ; 80%, 1 kHz Amplitude Modulation)
Frequency: 150kHz to 80MHz
PASS
VOLTAGE DIPS &
INTERRUPTIONS
IEC 60601-1-2
IEC
61000-4-11
0 % UT; 0,5 Cycle At 0°, 45°,90°, 135°,180°, 225°, 270°, 315°,
0 % UT; 1.0 Cycle
70 % UT; 25/30 Cycles
PASS
Template Doc.Id. Version
Core Text Template FL 09-04-01 TMP 03