Specifications

Document Doc.Id. Version
Name of document GR 21-005 01
16. Product Disposal
- Headset Pads are cellulose sponges and can safely be disposed of as household
waste.
- The headset and charger should be disposed of as electronic waste according to the
local legislation.
17. Contact Us
If you have any questions regarding Flow, don’t hesitate to contact customer support at
support@flowneuroscience.com
Flow Neuroscience AB
Baltzarsgatan 23; 211 36 Malmö; Skåne län, Sweden
18. EMC - Test Results
The device does not emit any radiation intended for
medical purposes
, neither hazardous nor
potentially hazardous radiation.
EMISSION LIMITS & MEASUREMENT
PHENOMENON STANDARD TEST METHOD TEST LEVELS RESULT
RADIATED RF EMISSION IEC 60601-1-2
CISPR 11
ETSI EN 301 489-1
30~1000 MHz ; 1~3 GHz & 3~6 GHz
(Measurement Uncertainty 4.24 dB)
PASS
CONDUCTED RF EMISSION IEC 60601-1-2
CISPR 11
ETSI EN 301 489-1
0,15~30 MHz
(Measurement Uncertainty 3.35 dB)
PASS
VOLTAGE
FLUCTUATION & FLICKER
IEC 60601-1-2
IEC 61000-3-3
EUT VALUES & LIMITS
Pst: 0.028 ; Limit 1.00
Plt: 0.028 ; Limit 0.65
dc [%]: 0.0000 ; Limit 3.30
dmax [%]: 0.053 ; Limit 4.00
dt [s]: 0.000 ; Limit 0.50
PASS
Template Doc.Id. Version
Core Text Template FL 09-04-01 TMP 03