Product Manual

FLIRCM72_CM74USERMANUALDocumentIdentifier:CM72_CM74enUS_AA
24
Table7.2Current(TRMS)
Function Range Accuracy
DCA(CM74only)
60.00A
±(2%+5dgt)
600.0A
ACA
60.00A ±(2%+5dgt)
45~65Hz
±(2%+5dgt)

65~400Hz
forCM74
±(2%+5dgt)
45~65Hz
±(3%+5dgt)
65~400Hz
forCM72
600.0A
ACAVFD
(CM74only)
60.00A
±(2%+5dgt)45~65Hz
±(6%+5dgt)65~400Hz
600.0A
Notes:
CM72:NoVFDModeandACAOnly
CM74:IntegratedVFDModealwaysONinACAMODE
Overloadprotection:600A(rms)
PositionError:±1%ofreadingforCM74(CM72notspecified).
ACConversionType andadditionalaccu racy issam easACVoltage.
DCAaffectedbythetemperatureandresidualmagn etism;use
DCAZerofunctiontocompensate.
ACconversiontype:ACcoupled,trueRMSresponding,calibratedtotheRMSvalueofasinewaveinput.
Accuraciesaregivenfor sinewavesatfullscaleandnonsinewavesbelowhalfscale.
Ifthemetermeasuresa4000countsignalandtheCrest
Factorofthesignalismorethen3.0,thereadingmay
notmeetspecifiedtolerances.Fornonsinewaves(50/60Hz),addthefollowingCrestFactorcorrections:
Fornonsinewaves(50/60Hz),addthefollowingcrestfactorcorrections:
Foracrestfactorof1.0–2.0,add3.0%totheaccuracy.
Foracrestfactorof2.0–2.5,add5.0%totheaccuracy.
Foracrestfactorof2.5–3.0,add7.0%totheaccuracy.