Manual
Table Of Contents
- 1 Disclaimers
- 2 Safety information
- 3 Notice to user
- 4 Customer help
- 5 Quick Start Guide
- 6 Description
- 7 Operation
- 7.1 Charging the battery
- 7.2 Turning on and turning off the camera
- 7.3 Saving an image
- 7.4 Recalling an image
- 7.5 Deleting an image
- 7.6 Deleting all images
- 7.7 Measuring a temperature using a spotmeter
- 7.8 Measuring the hottest temperature within an area
- 7.9 Measuring the coldest temperature within an area
- 7.10 Hiding measurement tools
- 7.11 Changing the color palette
- 7.12 Working with color alarms
- 7.13 Changing image mode
- 7.14 Changing the temperature scale mode
- 7.15 Changing the camera temperature range
- 7.16 Setting the emissivity as a surface property
- 7.17 Setting the emissivity as a custom material
- 7.18 Changing the emissivity as a custom value
- 7.19 Changing the reflected apparent temperature
- 7.20 Changing the distance between the object and the camera
- 7.21 Performing a non-uniformity correction (NUC)
- 7.22 Configuring Wi-Fi
- 7.23 Changing the settings
- 7.24 Updating the camera
- 8 Mechanical drawings
- 9 CE Declaration of conformity
- 10 Cleaning the camera
- 11 Application examples
- 12 Thermographic measurement techniques
- 13 About calibration
- 13.1 Introduction
- 13.2 Definition—what is calibration?
- 13.3 Camera calibration at FLIR Systems
- 13.4 The differences between a calibration performed by a user and that performed directly at FLIR Systems
- 13.5 Calibration, verification and adjustment
- 13.6 Non-uniformity correction
- 13.7 Thermal image adjustment (thermal tuning)
- 14 About FLIR Systems
Operation
7
4. To change the temperature span and the temperature level in Manual mode, do the
following:
• Push the navigation pad left/right to select (highlight) the maximum and/or mini-
mum temperature.
• Push the navigation pad up/down to change the value of the highlighted
temperature.
7.14.1 When to use Manual mode
7.14.1.1 Example 1
Here are two infrared images of a building. In the left image, which is auto-adjusted, the
large temperature span between the clear sky and the heated building makes a correct
analysis difficult. You can analyze the building in more detail if you change the tempera-
ture scale to values close to the temperature of the building.
Automatic Manual
7.14.1.2 Example 2
Here are two infrared images of an isolator in a power line. To make it easier to analyze
the temperature variations in the isolator, the temperature scale in the right image has
been changed to values close to the temperature of the isolator.
Automatic Manual
7.15 Changing the camera temperature range
The camera is calibrated for different temperature ranges. Available temperature range
options are dependent on the camera model.
For accurate temperature measurements, you must change the Camera temperature
range setting to suit the expected temperature of the object you are inspecting.
Note For more information, see section 13 About calibration, page 39.
#T559828; r. AQ/75691/75691; en-US
18