5 Operation
4. For stabilized voltage measurements, place a low-impedance load across
the probe inputs.
5. Connect the probe leads in parallel to the part under test.
6. Read the voltage value on the display.
5.5 Resistance measurements
WARNING
Do not do diode, resistance or continuity tests before you have removed the
power from the capacitors and from a device during a test. Injury to persons
can occur.
1. Set the function switch to the
position.
2. Ensure that the meter is set to resistance measurement. The Ω unit will be
displayed.
If the
or indicator is displayed, press the button repeatedly until
the Ω unit is displayed.
3. Insert the black probe lead into the negative
terminal and the red probe
lead into the positive
terminal.
4. Touch the tips of the probe across the circuit or component under test.
5. Read the resistance value on the display.
5.6 Continuity test
WARNING
Do not do diode, resistance or continuity tests before you have removed the
power from the capacitors and from a device during a test. Injury to persons
can occur.
1. Set the function switch to the
position.
2. Use the
button to select continuity measurement. The indicator will
be displayed.
3. Insert the black probe lead into the negative
terminal and the red probe
lead into the positive
terminal.
4. Touch the tips of the probe across the circuit or component under test.