User Manual

7 Technical specifications
Resolution 0.1%
Basic accuracy
±(5.0% + 10 digits) for order 1–12
±(10.0% + 10 digits) for order 13–
25
Ohms
Ranges 1.0000 kΩ, 10.00 kΩ, 100.00
Resolution 0.01 Ω
Accuracy ±( 1.0% + 3 digits)
Overload protection 1000 V RMS
Continuity beeper
<30 Ω, 2 kHz tone buzzer
Diode test
Open circuit voltage ±1.8 V max
Capacitance
Ranges 400.0 μF, 4.000 mF
Resolution 0.1 μF
Basic accuracy ±(1.9% + 8 digits)
Overload protection 1000 V RMS
Frequency counter
Ranges 20.0 Hz to 10 kHz
Resolution 0.1 Hz
Basic accuracy ±(0.5% + 3 digits)
Overload Protection AC/DC 600 A, AC/DC 1000 A
#T559825; r.8008/8011; en-US 28