8.6.2 Cleaning the thermal desorber
Occasionally, the thermal desorber (TD)can become dirty or contaminated. This will
adversely affect the performance of the Fido X4.
Materials needed
Alcohol swab or pad
Procedure
Step 1: Power off the Fido X4 and allow 15 minutes for surfaces to cool.
Step 2: Remove the TD by depressing the tab latches on each side.
Step 3: Using an alcohol swab or pad, clean the circular metal surface and the plastic sur-
rounding it.
Step 4: Re-install the TD and power the system on.
8.6.3 Cleaning a clogged or dirty sensor Inlet
Rarely, it is possible to clog or contaminate an inlet sufficiently to require more thorough
cleaning.
To clean the detector inlet:
Step 1: Power off the Fido X4 and allow to cool for 15 minutes.
Step 2: Use the inlet tool to remove the detector inlet tip as shown in the "Remove the
Inlet" video, available from the Help menu.
Step 3: Thread the inlet into the cleaning adapter Figure 8-11).
Figure 8-11: Cleaning adapter
Step 4: Fill the syringe with clean DI water or alcohol.
Step 5: Slowly depress the syringe to force liquid through the inlet.
Step 6: Discard solution appropriately.