Operations Manual

Table Of Contents
4226044 (Rev. C) © 2020 FLIR Systems, Inc.
19
FLIR Fido® X4 Operations Manual
3.5.1 Sensing element
Caution: To avoid burns, when handling the sensing element, take care only
to touch the end of the element that is marked with a dot. See Figure 3-4.
Caution: Breaking a sensing element will create sharp edges and glass shards
that can cut or puncture skin.
Do not store sensing elements in clothing pockets.
Do not put a sensing element where someone might sit or step on it.
Properly dispose of used or broken sensing elements with care to avoid
injury.
Figure 3-4: Sensing element (dot enlarged)
The Fido X4 sensing element contains the highly sensitive and selective fluorescent and
chemiluminescent sensing materials that make-up TrueTrace technology. During operation,
the unmarked end of the sensing element is inserted into the thermal desorber and heated.
The foundation of the sensing element is an optical-grade glass capillary tube. Sensing mater-
ials are applied as thin films, in a linear, 5-channel array, inside of the capillary. Multiple vari-
ations of these novel materials are used to achieve sensitivity and selectivity across a broad
range of threat materials, including:
nitroaromatics
nitrate esters
nitramines
nitrosamines
inorganic nitrate
nitro-volatiles
peroxides
Periodically, the sensing element must be replaced.During operation of the Fido X4, system
diagnostics continually monitor sensor array performance to determine when replacement is
needed.When the system determines that the sensing element must be replaced, simple
prompts are displayed to inform the user. For more information, see "Replacing the Sensing
Element" on p. 53.