Instruction Manual
FLINTEC
Document
Date
Code
Rev.
Page
FAD - 4 DIGITAL LOADCELL JUNCTION BOX
02.05.00
FAD4_2RM
2
11 of 20
Master transmission: (PAR 2)
STX ADD P 2 aa bb cc BCS2 BCS1 ETX
aa : Weight resolution (01, 02, 05, 10, 20, 50).
bb: The first 2 digits of weighing range (00-99).
cc: DUAL INTERVAL. The first 2 digits of the weighing range for which the lower weight
resolution will be selected automatically. 00 = DUAL INTERVAL NOT ACTIVE.
Slave response:
STX ADD ACK or NAK BCS2 BCS1 ETX
A NAK block will be returned if some parameter is out of range.
Master transmission: (PAR 3)
STX ADD P 3 a b c d e f BCS2 BCS1 ETX
a : Decimal digits (0, 1, 2, 3, 4)
b : Autozero maintenance 0 = NO 1 = YES
c : Autozero on power up 0 = NO 1 = YES
d : Dual digital filter 0 = NO 1 = YES (The filter doubles when not in motion)
e : Leading zero blank 0 = NO 1 = YES
f : Reserved : always 0
Slave response:
STX ADD ACK or NAK BCS2 BCS1 ETX
A NAK block will be returned if some parameter is out of range.
THE VALUE(S) ARE NOT SAVED IN EEPROM UNTIL A STORE COMMAND IS ISSUED.
p. UPLOAD PARAMETERS
Used to upload weight parameters.
The parameter groups a b c d e f are identical with the āPā command.
Master transmission: (PAR 1)
STX ADD p 1 BCS2 BCS1 ETX
Slave response:
STX ADD p 1 a b c d e f BCS2 BCS1 ETX
a : Conversions per sec : 1=8Hz/ n , 2=20Hz/ n , 3=40Hz/ n , 4=80Hz/ n. n=Nr of Loadcells.
MAXIMUM Conv/sec = 40Hz i.e. the 80Hz with one loadcell is not acceptable.
b : Gain set. 0 = 2 mV/V , 1 = m4 mV/V , for each input (Deadload + Span).
c : Number of loadcells (1, 2, 3, 4)
d : Weight units 0 = kg
e : Digital filter (0, 1, 2) 0 = 1 , 1 = 2 , 2 = 4 samples.
f : No motion samples (1, 2, 3, 4, 5, 6) consecutive conversions within 1e.
Master transmission: (PAR 2)
STX ADD p 2 BCS2 BCS1 ETX
Slave response:
STX ADD P 2 aa bb cc BCS2 BCS1 ETX
aa : Weight resolution (01, 02, 05, 10, 20, 50).
bb: The first 2 digits of weighing range (00-99).
cc: DUAL INTERVAL.The first 2 digits of the weighing range for which the lower weight
resolution will be selected automatically. 00 = DUAL INTERVAL NOT ACTIVE.