A/D Converter FAD-30 Technical Manual Flintec GmbH Bemannsbruch 9 74909 Meckesheim GERMANY www.flintec.com FAD-30 Technical Manual, Rev. 1.
Table of Contents: 1. SAFETY INSTRUCTIONS ...........................................................................................................................4 2. DECLARATION OF CONFORMITY ...........................................................................................................5 3. 3.1. 3.2. 3.3. 3.4. INTRODUCTION .........................................................................................................................................6 Overview ..................
10. FAD-30CO – WITH CANOPEN INTERFACE ...........................................................................................43 10.1. Front View ..............................................................................................................................................43 10.2. Electrical Connections ............................................................................................................................44 10.3. Interfacing.............................................
1. SAFETY INSTRUCTIONS CAUTION READ this manual BEFORE operating or servicing this equipment. FOLLOW these instructions carefully. SAVE this manual for future reference. DO NOT allow untrained personnel to operate, clean, inspect, maintain, service, or tamper with this equipment. ALWAYS DISCONNECT this equipment from the power source before cleaning or performing maintenance. CALL FLINTEC ENGINEERING for parts, information, and service. WARNING ONLY PERMIT QUALIFIED PERSONNEL TO SERVICE THIS EQUIPMENT.
2.
3. INTRODUCTION 3.1. Overview The type FAD-30 A/D Converter Series consists of powerful and economic state-of-the-art instruments for static and dynamic weighing applications plus force and torque measurements. Each instrument of the series converts the analogue low level signal from a load cell or a strain gauge sensor to a digital high-resolution and high-accuracy signal and transmits the digital data to an external PLC or PC system.
FAD-30PB FAD-30PN FAD-30EN FAD-30CO Bipolar signal input in Force and Count modes 500 to 100 000 counts external resolution at calibrated data Unipolar 1 million and bipolar 2 million counts is adjusted and filtered to external resolution in count mode High internal resolution up to 8 million counts Up to 800 conversions per second Standard Serial interface RS-232C Standard Serial interface RS-485A Modbus RTU interface Profibus DP-V1 interface ProfiNet interface Ethernet (Modbus TCP/IP or BSI) interface
Communication and Setup: Serial interface RS-232C: Other interfaces Response time Setup and calibration 9600 baud (8,N,1) Depends on instrument type < 4 ms (delay after each read or write command) By PC software via RS232C, backup data stored on PC Power supply: DC power supply 10 to 28 VDC, < 200 mA, not galvanically isolated Environment and Enclosure: Operation temperature: Enclosure and protection Between -10 °C and +40 °C at maximum 85% RH max, non-condensing Polyamide, for DIN-rail mounting, IP20
3.3. Housing FAD-30 instruments come within a polyamide housing sealed to IP20. They are prepared for mounting on NS 37/7 or NS 35/15 standard DIN rails (see drawings). 45 mm 99 mm View A View B Side view 114,5 mm 22,5 mm View A: FAD-30, FAD-30MB View B: FAD-30PB, FAD-30PN, FAD-30EN, FAD-30CO Figure 3.2 – Dimensions 3.4.
4. INSTALLATION PRECAUTION: Please read this manual carefully before you install the instrument. If you apply all recommendations in this chapter you will increase the reliability and long term performance of your system. 4.1. Recommendations 4.1.1. Control Cabinet Design Warning: Please follow the following warnings for designing the control cabinet which will increase the reliability of your system. The control cabinet should be designed therefor the A/D converters can operate safely.
The quality of the instrument‟s ground will determine the accuracy and the safety of your measuring system. A proper ground connection is needed to minimize extraneous electrical noise effects on the measurement. A poor ground can result in an unsafe and unstable operation. It is important that the instrument should not share power lines with noise-generating equipment such as heavy load switching, motor control equipments, inductive loads, etc.
5. SETUP PRECAUTION: Please read this manual carefully before energizing the instrument. Perform the commissioning according the procedure given in chapter 4.3. Only trained person are allowed for commissioning, checking, cleaning and servicing of the instrument. The interference of untrained person may cause some unwanted damages or injures. FAD-30 series instruments are setup and calibrated by the xFace software supplied with the instrument.
5.3. A/D Converter Setup and Calibration WARNING: Read this manual carefully before you setup and calibrate the instrument. This will increase the performance of your weighing system by applying proper setup and calibration. Visual Weight Display: Displays the weight, count or force value Toolbox: Contains shortcuts of some special commands A/D Converter Parameters: This block allows the user to setup the A/D Converter‟s parameter related to the operation mode. Refer to chapter 5.3.
Motion Detection This parameter defines the motion detection window which determines a stable weighing. If motion detection is not required, this parameter can be disabled. The available motion detection window values are: Weighing and Force Mode: Disable ± 0.3e ± 0.5e (default setting) ± 1e ± 2e ± 60 ± 100 (default setting) ± 200 ± 400 Count Mode: Disable Power On Zero This parameter enables automatic zeroing after powering on the instrument.
Checksum The checksum can be enabled or disabled within BSI and Continuous output data format. The checksum calculation can be found in the related data format description. Default setting: „Enable‟. Carriage Return Carriage Return can be enabled or disabled within Continuous output data format. Default setting: „Enable‟. Line Feed Line Feed can be enabled or disabled within Continuous output data format. Default setting: „Enable‟.
Mode selection FAD-30 instruments have three operation modes which are: Count Mode: Weighing Mode: Force Mode: Filtered and normalized ADC count data will be transmitted in this mode. The calibration is performed at the PLC, if any. Unipolar weight data in calibrated weighing unit will be transmitted. This mode is compatible to OIML R76 and EN 45501. Bipolar or unipolar force measurement data in calibrated force unit will be transmitted. Select the scale operation mode and the polarity.
Zero Adjustment If your scale has a residual zero drift you may perform a zero adjustment only. Select “Zero Adjustment” and press the “Start” button. Then the zero calibration window will be displayed. Unload the scale and click the “OK” button. The [WAIT] message appears on the virtual display during the zero adjustment. Approximately 5 seconds later the zero adjustment will be finalized. Note: Zero adjustment is also performed over the bus interface.
5.4. A/D Converter Performance Test The scale performance test should be performed before you install the bus connection. The tests are linearity, repeatability and eccentricity. You can follow the test at the virtual display and/or status tab which show the measurement data, the instrument data and the software version. For convenience keys for Taring, Zeroing and Clear are located in this tab. 5.5.
5.8. Continuous Output Data Structure (for FAD-30 + FAD-30MB only) The continuous data output of the instrument will be transmitted in the following data structure at the speed which is determined by the baud rate and the data length. The definition of the continuous data format is described below.
5.9. BSI Data Structure (for FAD-30 + FAD-30MB only) Depending on their functionality all new generation Flintec instruments launched on the market support the standardized command set with BSI data structure. This easy data format gives a reliable and fast interface for communicating with PLC or PC for process control or transactional applications. You can expand your system with additional scales from Flintec without having to change your application program base. General Rules: 1. 2. 3. 4.
Commands and Responses A Command: Response: Example: Read all weight data [ADR][A] [ADR][A][STATUS][SIGN][NET W][SIGN][TARE W][SIGN][GROSS W] Command : Response : Comments: B Command: Response: Example: 01A 01AS+000123.4+000111.1+000234.5 01AD+000123.4+000111.1+000234.5 01AO ( ADC out error ) The response is net, tare and gross weight values or error status. All weight data will be transmitted immediately after receiving the command.
S Command: Response: Example: Read Status [ADR][S] [ADR][S][STATUS-1][STATUS-2][STATUS-3] Command: Response: Comments: T Command: Response: Comments: X Command: Response: Example: 01S 01SSGI (Stable, Gross, In Range) 01SDGL (Unstable, Gross, Low voltage error) The response includes 3 status information. STATUS-1 can be Stable or Unstable. STATUS-2 can be Gross or Net. STATUS-3 can be „in range‟, „out of range‟, „low voltage‟ or „high voltage‟.
5.10.
Address R/W Word Command 40015 R/W 1 mV operation in Count Mode 40016 R/W 1 Digital filters Definition 0 1 2 3 0 1 2 3 4 5 6 7 8 9 5 mV 10 mV 15 mV 18 mV Fast Medium Default Slow EXPLANATION Attention: For hardware connection details, please refer to the related hardware descriptions in this manual. In the two word registers the data is stored to the registers in big-endian format.
5.11.
Calibration Status nd Bit no. D31…D16 D15…D8 D7...D0 2 Dword descriptions when read command is ‘Calibration Status’.
PLC Output to FAD-30Px Input Dword D31 D15 D30 D14 D29 D13 D28 D27 D12 D11 D26 D10 D25 D9 st 1 Dword D24 D8 D23 D7 D22 D6 D21 D5 D20 D4 D19 D3 D18 D2 D17 D1 D16 D0 Not in use W31 W30 W29 W28 W27 W26 W25 W24 W23 W22 W21 W20 W19 W18 W17 W16 MSB 2 Dword W0 W15 W14 W13 W12 W11 W10 W9 W8 W7 W6 W5 W4 W3 W2 W1 LSB rd 3 Dword Not in use New Not in use Command List Data Selection CMD nd PLC Output to FAD-30Px Input 2nd Dword nd The 2 Dword contains the setting / parameter value (Refer to “Comman
5.12.
Calibration Status Bit no. D31…D16 T_DW2 descriptions when read command is ‘Calibration Status’.
PLC Output to FAD-30CO Input Dword D31 D15 D30 D14 D29 D13 D28 D27 D12 D11 D26 D10 D25 D9 RxPDO 1 (R_DW1) D24 D8 D23 D7 D22 D6 D21 D5 D20 D4 D19 D3 D18 D2 D17 D1 D16 D0 Reserved W31 W30 W29 W28 W27 W26 W25 W24 W23 W22 W21 W20 W19 W18 W17 W16 RxPDO 2 MSB (R_DW2) W0 W15 W14 W13 W12 W11 W10 W9 W8 W7 W6 W5 W4 W3 W2 W1 LSB Not in use RxPDO 3 New (R_DW3) Not in use Command List Data Selection CMD PLC Output to FAD-30CO Input R_DW 2 R_DW 2 contains the setting / parameter value (Refer to “Adjust
6. FAD-30 / FAD-30MB – WITH RS-485 INTERFACE FAD-30 and FAD-30MB instruments are state-of-the-art strain gauge load cell signal digitizers with RS-485 interface. These instruments are used for any type of process weighing and force measurement including tank and silo weighing, dynamic weighing, check weighing, filling, tension /compression force measurement etc. 6.1. Front View There are 3 status LEDs on the front panel which indicate the operational instrument status (Refer to table 6.1).
6.2. Electrical Connections RS-232 and RS-485 connections are shown in figure 6.2. Figure 6.
6.5. Modbus RTU Setup (for FAD-30MB only) If the FAD-30MB instrument is setup for Modbus, it can be used as a Modbus RTU slave in a RS-485 communication network. Function codes „0x03‟ and „0x10‟ are supported. For RS-485 setup please refer to chapter 5.3.2, for the Modbus data structure please refer to chapter 5.10. FAD-30 Technical Manual, Rev. 1.
7. FAD-30PB – WITH PROFIBUS INTERFACE FAD-30PB instruments are state-of-the-art strain gauge load cell signal digitizers with Profibus DP interface. These instruments are used for any type of process weighing and force measurement including tank and silo weighing, dynamic weighing, check weighing, filling, tension /compression force measurement etc. The GSD file is available on a CD which is supplied together with the instrument. 7.1.
ST Status LED State Off Green Flashing Green Red Indication No power or not initialized Initialized Initialized, diagnostic event(s) present Exception error Comment No power or profibus module is in initializing Diagnostic is active There is an exception error OP Operation Mode LED State Off Green Flashing Green Flashing Red (2x) Indication Not on-line / No power On-line, data exchange On-line, clear PROFIBUS configuration error Comment Check power and cable Check GSD file configuration. 7.2.
Load Cell Connection See chapter 4.2.2 Power Supply Connection: See chapter 4.2.1 7.3. Interfacing FAD-30PB instruments have two interface connectors on the front of the instrument: Profibus and RS-232C. The table below describes the data formats for interfacing the peripherals and their application. Data Format Profibus Description Profibus interfacing. Refer to chapter 5.11 Application Interfacing with PLC Hardware Profibus-DP-V1 Table 7.
8. FAD-30PN – WITH PROFINET INTERFACE FAD-30PN instruments are state-of-the-art strain gauge load cell signal digitizers with ProfiNet interface. These instruments are used for any type of process weighing and force measurement including tank and silo weighing, dynamic weighing, check weighing, filling, tension /compression force measurement etc. The EDS file is available on a CD which is supplied together with the instrument. 8.1.
MS Module Status LED LED State Description Off No power or not initialized Green Green, 1 flash Initialized Initialized, diagnostic event(s) present Green, 2 flashes Blink Red Red, 1 flash Red, 2 flashes Red, 3 flashes Exception error Configuration error IP address error Station name error Red, 4 flashes Internal module error Comment No power or ProfiNet module is in initialization state Diagnostic is active Used by engineering tools to identify the node on the network There is an exception erro
Load Cell Connection See chapter 4.4.2 Power Supply Connection See chapter 4.2.1 8.3. Interfacing FAD-30PN instruments have two interface connectors on the front of the instrument: ProfiNet and RS-232C. The table below describes the data formats for interfacing the peripherals and their application. Data Format ProfiNet Description ProfiNet interfacing. Refer to chapter 5.11 Application Interfacing with PLC. Hardware ProfiNet Table 8.
9. FAD-30EN – WITH ETHERNET INTERFACE FAD-30EN instruments are state-of-the-art strain gauge load cell signal digitizers with Ethernet interface. These instruments are used for any type of process weighing and force measurement including tank and silo weighing, dynamic weighing, check weighing, filling, tension /compression force measurement etc. 9.1. Front View There are 3 status LEDs on the front panel which indicate the operational instrument status (Refer to table 9.1).
9.2. Electrical Connections Ethernet and RS-232C connections are shown in figure 9.2 to 9.4.
Load Cell Connection See chapter 4.2.2 Power Supply Connection See chapter 4.2.1 9.3. Interfacing FAD-30EN instruments have two interface connectors on the front of the instrument: Ethernet and RS-232C. The table below describes the data formats for interfacing the peripherals and their application: Data Format Modbus TCP/IP Description Ethernet interfacing. Refer to chapter 5.10 Application Interfacing with PLC Hardware Ethernet Table 9.
10. FAD-30CO – WITH CANOPEN INTERFACE FAD-30CO instruments are state-of-the-art strain gauge load cell signal digitizers with CANopen interface. These instruments are used for any type of process weighing and force measurement including tank and silo weighing, dynamic weighing, check weighing, filling, tension /compression force measurement etc. The EDS file is available on a CD which is supplied together with the instrument. 10.1.
RUN LED State Off Green Green, flashing Green, single flash Green, flickering Red Indication No power or not initialized On-line, data exchange On-line, initializing Stopped Auto baud rate detection in progress Exception error Comment No power or cable broken Check for hardware damages Check EDS file ERR LED State Off Red, single flash Red, flickering Indication No error Warning limit reached LSS Red, double flash Error count event Red Bus off (fatal event) Comment No power or CANopen module is in
CANopen Interface Use Data format Interfacing with PC or PLC CANopen Automatically detected and supported baud rates are 10 kbps, 20 kbps, 50 kbps, 100 kbps, 125 kbps, 250 kbps, 500 kbps, 800 kbps, 1 Mbps; Autobaud is default setting. Baud rate Warning: Connect the shield to the reference ground or shield pin of the power connector. Warning: Disconnect xFace PC software for Profibus interfacing. Load Cell Connection See chapter 4.2.2 Power Supply Connection: See chapter 4.2.1 10.3.
11. ERROR TABLE The FAD-30 A/D Converters have been designed as very reliable and virtually error free instruments. However if an error occurs do not attempt to repair the equipment before you understand what caused the error. Note the problems you have with your instrument and the error messages shown by the LEDs located on the front panel. Then try to solve the problem according to the error table given below.
12. DIAGNOSTICS In this test menu serial interface tests (RC-232C and/or RS-485) and load cell signal analog to digital conversion and processing tests are performed sequentially. For entering the diagnostics mode, press the setup switch before power on and release the switch after the instrument is powered on. The instrument will go into the RS-232C RXD test mode which is indicated by lighted LED and flashed Err LED as shown below.
13. FREQUENTLY ASKED QUESTIONS Question : Answer : Question : Answer : Question : Answer : Question : Answer : Question : My PC has a COM port but I cannot see any COM port in the Connection Setting menu. How can I solve that problem? Answer : Another software may be connected to that COM port. Close all applications before you start xFace. Question : Answer : My PC cannot interface with FAD-30. How can I check the COM ports? Short circuit the RXD and TXD pins of your COM port.