Product Overview
22
£
WARNING: Before making a pilot flame test, manually shut off the fuel supply to the main
burner.
Normal Pilot Flame Test (MP560, MP561, MP562 Programmers Only)
1. Place the “Run-Check” switch in the “Check” position.
2. Turn power on and initiate a normal startup.
3. Observe the pilot flame signal on the test meter. If the average flame is below normal, a steady
DC voltage reading of 4.0 to 6.0 volts (for UV amplifiers) and 14 to 18 volts (for flame rectifica-
tion amplifiers), re-adjust the pilot flame or realign the flame detector.
WARNING: DO NOT TOUCH a flame rectification rod with power applied.
4. During the pilot flame test and adjustment period, if flame is not detected within 30 seconds, the
control will lock out. To reestablish the pilot flame trial for ignition (P.T.F.I.), manual reset of the
lockout switch is required, and a complete repurge is accomplished.
5. When UV detection is used, a test is required to verify that UV radiation from the ignition spark
is not being detected. To accomplish this, manually shut off both pilot and main fuels. Initiate a
normal startup, observe the test meter which should read no more than 1/2 volt DC. If more than
1/2 volt DC is observed, realign the UV scanner, and/or shield the spark from the scanner’s
view.
6. Move the “Run-Check” switch to the “Run” position, check pilot flame failure response time by
manually shutting off the pilot fuel and then initiate a normal startup. With no pilot flame
present, the control will de-energize the pilot assembly at the end of the trial for ignition interval
(5 or 10 seconds, selected by dipswitches #6 and #7 - see Programmer Dipswitch Settings on
page 11), and the control will lock out.
WARNING: The minimum pilot test must be accomplished by a trained and qualified burner
technician.
Minimum Pilot Test
This test insures that the flame detector will not sense a pilot flame too small to light the main flame
reliably. It must be made on every new installation as well as following the repositioning of the flame
detector. This procedure should not be used on a direct spark ignited burner.
1. Manually shut off the fuel to the main burner.
2. Place the “Run-Check” switch in the “Check” position. (MP560 Programmers only).
3. Connect a test meter to the test jacks on the Amplifier Module.
4. Initiate a normal startup.
5. Reduce the fuel to the pilot until the DC voltmeter reads 3.5 volts for UV scanners. See WARN-
ING below. This is the minimum pilot. For flame rectification the flame signal for minimum
pilot varies depending on the application. See WARNING below.
6. Return the “Check-Run” switch to the “Run” position. (MP560 Programmer only).
7. Slowly turn on the main fuel and insure that the main flame lights off promptly and normally.
WARNING: If light off is delayed, shut off the power to the installation. Realign the flame
detector so that pilot flame detection requires a larger pilot flame. Repeat this test until the
main flame lights reliably with minimum pilot.