Datasheet

9. TIME DEPENDENCY CHARACTERISTICS
Fig:ll
shows
a.typical
Pattern
for a TGS
#813
Sensor
which has been stored
unenergized for a long period
of.time.
This graph
Shows
Sensor
resistance
at a
constant
gas concentration.
When the
Sensor
is
energized after a
lang
storage period it takes
approx-
imately 1-10 days for the
Sensor
to
resch
a stable resistance level.
Please note,
that the
actual
transitional time necessary to
resch
stability
is
dependent on the amount of time the
Sensor
has been stored and the
atomospheric conditions
under
which it has been stored.
.'
Table
ID
suggested that the
Sensor
be conditioned for at least 7 days
Prior
to testing and evaluation.
This period is necessary
because
of the phenomenon
illustrated in
Fig.11.
.
L
Caution at Calibration Time
.
Regardingtheabove characteristics
of the
Sensor,
we strongly recommend that
you wait until the
Sensor
has reached a stable output level before you
calibrate.
Failure to do so may result in
non-Standard
Sensor
perfromance.
I
@
Initial
and
transitional period
a?ter
switch
on.
@
Stabillzed
period of energited output level.
&
1.0
-.-
a
Ib:
.+
,.,.
*,
:,
/
m
1 hour after
restoration
of
current
input
1-l
Odays after
restoration of
cunent input
Time
FIG.
11.
TYPICAL PATTERN OF TGS $813
SENSITIVITY
CHANGE WITH TIME.
+Remarks;
Ro: Sensor resistance at
stabiiited
level.
Fl:
Sensor resistance during s?abilization.
(13)