User Manual
3534
Delete Log File
Clear up space by deleting old logs or just view
free space available.
MENU/Delete Log File
1. Use ARROWS to scroll through Jobs (logs). The display will alternate
between the % free space on the drive and the % of space the chosen
Job uses.
2. Press ENTER to select a Job to delete. Choose whether or not you
want to delete that Job.
3. If you choose to delete, it may take a few seconds to finish. If no
more Jobs are found, the manifold returns to standard operation.
• Press MENU to exit at any time.
Format Internal Flash Drive
Quickly clear up maximum space by
reformatting the internal flash drive. This deletes
everything on the drive including log files,
firmware update files, and any other files added
manually.
MENU/Format Drive
1. Use ARROWS to select Yes or No.
2. Press ENTER to select and exit.
• Press MENU to exit at any time.
• If you choose to format, it may take a few seconds before returning
to standard operation.
• User settings are not deleted.
Maintenance
Cleaning
Wipe with damp cloth to clean the exterior.
Do not use solvents.
Battery Replacement
The batteries must be replaced when the
battery life indicator is empty. Once the batteries
are drained beyond operating voltage, “Low Bat”
briefly appears and the manifold turns off.
Unscrew the 4 cover screws and pull out the
rear battery cover. Replace the 6 AA batteries and
properly dispose of the old ones.
Using Different Refrigerants
You can use different refrigerants, but be
sure to purge your manifold with nitrogen
before connecting to a system with a different
refrigerant. Contamination can hurt system
performance and cause damage.
Temperature Calibration
See page 25.
Advanced Pressure Calibration
See page 30.