User's Manual

Getting started
24
Site requirements
z
Ensure that the laboratory floor is level and the laboratory table has a
weight capacity of over 200 kg (441 lb).
z
Ensure that the laboratory is free of dust, corrosive and flammable gas, and
heat and wind sources.
z
Ensure that the laboratory is away from direct sunlight and well ventilated.
We recommend that you refer to the standard of a biosafety level (BSL) 2
laboratory.
z
The device should be used with the ultra-pure water machine and
refrigerator, and can be used with the genetic sequencer. Therefore,
enough space should be considered when placing the above instruments.
z
Ensure that enough space is provided around the device for ventilation,
cable connection, and power switch operation.
The following figure indicates distances that are required for optimal
operation and access.
200 mm
(8 inches)
200 mm
(8 inches)
200 mm
(8 inches)
Device
Operation area
Figure 8 Space requirements
Network requirements
Operating environments of the computer
Minimum configuration
Processor: Intel Celeron CPU J1900