User's Manual

Device overview
18
Control software
The software of the device can guide the user to load different
sample libraries and/or reagents to a sequencing flow cell according
to experimental requirements.
The following table describes the function of each functional
module:
Item Description
Self-test
Checks whether the components of the system are
functional.
Loading
Loads the required sample library and/or reagent to the
flow cell.
Wash
Performs wash and maintenance for the fluidics lines of
the system.
Self-test interface
After you power on the device, self-test starts. If the self-test
succeeds, the main interface appears.
If the self-test fails, perform the following steps:
1. In the main interface, tap , and select Log to check the
detailed self-test results that are recorded in the log.
2. Follow the on-screen instructions or the solutions that are
mentioned in
Troubleshooting on Page 41
.
3. Perform self-test again:
¾
Tap , select Maintenance > Self-test.
¾
Tap , select Shut down > Restart.
If the problems still persist, contact the technical support.