Users Manual Part 2
Reviewing parameters
43
6. Tap Advanced settings and you can enable Custom primers and
Auto wash if needed.
In the sequencing process, you can tap to enable or disable
Auto wash if needed.
7. Tap Next.
If any error prompts occur, follow the on-screen prompt to solve
the problem and tap Next again.
Reviewing parameters
Carefully check each item in the review interface, and do one of the
following:
z
If you find any error, tap Previous to return to the previous
interface and reset parameters.
z
If all parameters are correct, tap Start. Ensure that the pure water
volumn is sufficient and tap Continue. The software automatically
checks whether system modules are working properly.
¾
If all modules are working properly, the system automatically
starts a sequencing run.
¾
If the system prompts that the storage space is insufficient,
solve the problem and tap Start to check again.
If other problem occurs, contact the technical support.
Performing a sequencing run
Warning
z
Ensure that all compartment doors are closed. Sequencing run cannot be
started when the reagent compartment door is open.
z
Only open the reagent compartment door when necessary to avoid adverse
effects on sequencing results or even damage to the device.
z
Do not bump, move, vibrate, or impact the device during sequencing.
Otherwise, sequencing results might be inaccurate.
z
Pay special attention to the LED status bar, icons, and prompts. If
errors occur, a message appears on the screen. Follow the prompt to
troubleshoot the problem. For information about the troubleshooting, refer
to
Troubleshooting on Page 52.