Users Manual Part 2
Maintaining the pure water container
51
¾
Tap and a confirmation dialog box appears. Select Yes to
stop wash.
5. After wash, tap Finish to return to the main interface.
6. Remove the washing flow cell, sequencing cartridge, and washing
cartridge.
7. Store the washing flow cell at room temperature.
The washing flow cell can be reused for only one month. Dispose
of the flow cell according to local regulations and safety standards
of your laboratory.
8. Empty the remaining solution in the washing cartridge into an
appropriate waste container.
9. Dispose of the washing cartridge according to local regulations and
safety standards of your laboratory.
Maintaining the pure water
container
Perform the following maintenance every week:
1. Empty the pure water container.
2. Spray 75% alcohol onto the inner surface of the lid and the surface
of the pure water tube, and then wipe them with a clean cloth.
3. Add sufficient laboratory-grade water into the pure water
container, and attach the lid back onto the container.
4. Gently swirl it until all inner walls are cleaned.
5. Empty the pure water container.
6. Repeat steps 3 to 5 twice.
Replacing the waste container
Chemicals in reagents and waste might cause personal injury through
contact with the skin, eyes, and mucosa. Wear protective equipment (such
as laboratory coat, and disposable bouffant cap, protective glasses, mask,
gloves, and shoe covers) when handling the waste. Personal injury might be
caused by contact with the skin, eyes, and mucosa.