Users Manual Part 2
Maintaining the power supply
49
Monthly cleaning
Warning
z
We do not recommend that you use other disinfectants or wash solutions
except for those that are mentioned in this manual. Because other solutions
are not verified for use and their effects to the device are unknown.
z
If you have questions about the compatibility of wash solutions, contact the
technical support.
Perform the following steps:
1. Power off the device and open the reagent compartment.
For details about powering off the device, refer to
Powering off
the device on Page 32.
2. Clean the surface of the device with a 75% alcohol wipe. Ensure
that the surface is free of DNBs, reagents, blood, and saliva.
Maintaining the power supply
z
When the device is not in use for seven days or longer, perform a
wash manually according to the relevant reagent kit user manual.
Turn off the device after the wash.
For details about powering off the device, refer to
Powering off
the device on Page 32.
z
Check whether the power cord and cables are in good condition
regularly. Contact the technical support if new cables are required.
Maintaining the software
After the receival of
Release Notification
, contact the technical
support to update the software.