6.1.2 RS-232C Output Data
Result data is output every time the operation is finished.
The result output is possible for each unit that has a SAN –DP2S/SP4S installed.
Use SAN –IF unit if you want to output multiple results.
Result Output Data
Data Length: 78 bytes fixed (78 letters)
Normal data format
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
COUNT Number Press
Number
PARM
Number
30H 30H 30H 31H 20H 20H 20H 20H 30H 31H 20H 20H 20H 31H 20H 20H
0 0 0 1 0 1 1
17 18 19 20 21 22 23 24 25 26 27 28 29 30 31 32
PEAK LOAD Judg
ment
FINAL DISTANCE Judg
ment
31H 32H 2EH 33H 34H 20H 20H 20H 31H 32H 2EH 33H 34H 20H 20H 20H
1 2 . 3 4 1 2 . 3 4
33 34 35 36 37 38 39 40 41 42 43 44 45 46 47 48
FINAL LOAD 1ST RATE Judg
ment
31H 32H 2EH 33H 34H 20H 20H 20H 30H 2EH 31H 32H 33H 20H 20H 20H
1 2 . 3 4 0 . 1 2 3
49 50 51 52 53 54 55 56 57 58 59 60 61 62 63 64
2ND RATE Judg
ment
1ST TIME Judg
ment
30H 2EH 31H 32H 33H 20H 20H 20H 20H 31H 30H 2EH 30H 20H 20H 20H
0 . 1 2 3 1 0 . 0
65 66 67 68 69 70 71 72 73 74 75 76 77 78
2ND TIME Judg
ment
Judg
ment
CR LF
20H 20H 32H 2EH 30H 20H 20H 20H 4FH 20H 20H 20H 0DH 0AH
2 . 0 O
Overall Judgment
"O":4FH ACCEPT
"X":58H REJECT
"0001 01 1 12.34 12.34 12.34 0.123 0.123 10.0 2.0 O" +CR+LF
Judgment letters displayed Judgment letters displayed Judgment letters displayed
beside PEAK LOAD value: beside FINAL DISTANCE value: beside 1
ST
& 2
ND
RATE value:
“ ” (20H): ACCEPT “ “ (20H): ACCEPT “ “ (20H): ACCEPT
“H” (48H): Above High Value “H” (48H): Above High Value “H” (48H): Above High Value
“L” (4CH): Below Low Value “L” (4CH): Below Low Value “L” (4CH): Below Low Value
“A” (41H): Abnormal Judgment letters displayed
“S” (53H): STOP beside 1
ST
RATE value:
“B” (42H): BYPASS “Z” (5AH): 1
ST
Zone Reject
Judgment letter displayed beside TIME value:
“ “ (20H): ACCEPT “H” (48H): Above High Value
COUNT Number: ‘0001’ is displayed after powering ON. 1 is added after each cycle.