User guide


9.3 Controller Error
WARNING:
DO NOT CONNECT OR DISCONNECT CABLES OR OTHER SYSTEM
COMPONENTS WITH POWER APPLIED.
9.3.1 Spindle Dual Port RAM Abnormal
<ERROR DESCRIPTION>
An Axis Unit Dual Port RAM Error occurred while attempting communications between the
Axis Unit and the Main Unit.
<CHECK ITEMS>
1. Check the connected Axis Units for indications of abnormality. A locked-up display, an
illuminated Abnormal LED, and failure to count down are all possible indicators of a
malfunctioning Axis Unit. If all Axis Units appear to be functional, the problem is likely to
be found in another System component. Use the Axis Unit (if applicable) connected
directly to the Main Unit ribbon cable to perform the following inspections.
2. Turn off the power and verify the communication bus ribbon cable is in good condition and
is properly connected to both the Main Unit and the Axis Unit.
3. Remove the Main Unit Multi PCB, and verify the software is compatible with the Axis Unit
PCB. This may be of concern if a component has been replaced.
4. Verify correct installation of all removable IC chips on both the Axis Unit and Multi PCB.
5. Reinstall the Multi and Axis Unit PCBs, as necessary. Ensure that the boards are properly
seated.
6. Ensure that the Main Unit and Axis Unit are not located near any strong electrical or
magnetic fields.
7. Verify correct input voltages and ensure the power cables are properly connected.
8. Turn on the power. If problem persists, replace the Multi PCB. (Refer to Section 8.3)
9. If problem still exists, replace the affected AXIS UNit (if applicable). (Refer to Section 8.3)
10. If problem still exists, replace the entire Main Unit. (Refer to Section 8.3)
9.3.2 Preamp EEPROM Read Abnormal
<ERROR DESCRIPTION>
Data error detected in the tool EEPROM information.
<CHECK ITEMS>
1. Check the connected Axis Units for indications of abnormality. A locked-up display, an
illuminated Abnormal LED, and failure to count down are all possible indicators of a
malfunctioning Axis Unit. If all Axis Units appear to be functional, the problem is likely to
be found in another System component. Use the Axis Unit (if applicable) connected
directly to the Main Unit ribbon cable to perform the following inspections.
2. Turn off the power and verify the communication bus ribbon cable is in good condition and
is properly connected to both the Main Unit and the Axis Unit.
3. Verify the homerun cable is in good condition and is properly connected to both the Axis
Unit and the tool (preamp / sensor).
4. Remove the Main Unit Multi PCB, and verify the software is compatible with the Axis Unit
PCB. This may be of concern if a component has been replaced.
5. Verify correct installation of all removable IC chips on both the Axis Unit and Multi PCB.
Chapter 9: Troubleshooting User Console Version A3.53
Page 9-13