Oil & Gas Wireless Enabled Monitoring System 39
Installer’s Guide
iTLM-1-Specific Menus
The iTLM-1 tabs include menus for calibrating the temperature and tank level
sensors on the iTLM-1 device. Figure 3-23 below shows the menu for calibrating
the temperature sensor. Figures 3-22 and 3-24 on the following pages show the
distance and depth calibration menus, respectively, for the tank level sensor. Luna
iMonitoring performs the temperature and distance calibrations for the
temperature and tank level sensors, respectively, before shipping. The installer
performs the depth calibration for the tank level sensor.
iTLM-1 Temperature Calibration
To set the temperature calibration parameters, select the Temp Cal tab at the
bottom of the screen to access the temperature calibration menu shown in
Figure 3-22 below. Enter the correct Slope and Intercept, and hit the Set
Calibration button. The iTLM-1 uses the temperature data to compensate for
the effects of temperature on the tank level measurement.