User's Manual

WEMS Installer’s Guide 35
WEMS Software
iTLM-1-Specific Menus
The iTLM-1 tabs include menus for calibrating the temperature and tank level
sensors on the iTLM-1 device. Figure 24 below shows the menu for calibrating
the temperature sensor. Figure 25 and Figure 26 on the following pages show
the distance and depth calibration menus, respectively, for the tank level sensor.
Luna iMonitoring performs the temperature and distance calibrations for the
temperature and tank level sensors, respectively, before shipping. The installer
performs the depth calibration for the tank level sensor.
iTLM-1 Temperature Calibration
To set the temperature calibration parameters, select the Temp Cal tab at the
bottom of the screen to access the temperature calibration menu shown in
Figure 24 below. Enter the correct Slope and Intercept, and hit the Set
Calibration button. The iTLM-1 uses the temperature data to compensate for
the effects of temperature on the tank level measurement.
Figure 24.
i
TLM-1 Temperature Calibration Menu.