Medical Compressed Gas Connectors
MAINTENANCE
Maintenance Checklist
Daily
Inspect for Leak-tight seal
The main seal must be replaced more frequently depending on wear. Dismantling of the connector
for this purpose is not required. It is recommended that an O-ring pick be used for removal to avoid
damage to the groove. Clean groove if required and insert new O-ring.
Connection should operate smoothly. If the connector is forced, remove from service.
Check for contamination, bent or missing components.
Check for leaks.
Weekly
Inspect for correct function
Inspect the correct engagement of the valve, collets or ttings.
Check for any bent or missing components.
Periodic
Inspect that all threaded components are tight and properly torqued.
Check for any bent or missing components.
Check for proper actuation of handle, collets and all moving components.
Check for leaks.
Medical and Industrial CGA 540
Figure 29. Check retaining ring to make sure it is tightly
seated in its groove.
z Check internal thread collets for a xed-center position
and even spacing.
z An “out-of-round” condition may hinder sleeve operation.
A visual inspection of the sleeve is usually sufcient.
Retaining
Ring