Upper Arm Blood Pressure Monitor Model :FDBP-A10BL USER MANUAL Subject to our actual product. ■ Thanks for using our product. ■ Take care of this instruction manual. ■ Please read this manual carefully before using. Precaution: Please read the instruction manual carefully before use it. If you have any questions, please consult the salesperson in the purchasing center or ask your doctor.
CONTENT 1. Quick instructions ........................................................................1 2. Product Structure Description ...................................................1 3. Warnings and precautions ..........................................................2 4. Instruction ........................................................................................2 5. Product operation instructions ................................................3 6. Connect the cuff ........................
1.Quick instructions 2.Product Structure Description 2.1 Upper Arm Blood Pressure Monitor 2.
2.3cuff 3. Warnings and precautions 3.1 Warning • This product is not suitable for newborns. • This product is for home use only. When the measurement is complete, consult your doctor if you have any doubts about the measurement results.Do not self-diagnosis and treatment according to measurement results. • When common arrhythmia (such as atria premature beats, premature ventricular and atrial fibrillation) occurs, it will affect the measurement accuracy of blood pressure.
3.2 Precautions • Avoid high temperature, humidity, dust and direct sunlight. • Avoid damage to the arm band and rubber tube due to folding • Do not disassemble this unit yourself! Otherwise it will cause the manufacturer's school pressure calibration to fail! • Avoid falling or violently vibrating the body. • The normal measurement period is about 1 minute. If the measurement is not stopped for a long time, please press the [START/STOP] button to stop the measurement.
Unpack the box and lay the product flat on the table. Note:If you bring the product from a storage environment, place it in a normal use environment for 20 minutes 5.2 battery If there is a signal of low power on the display, connect your charger. Icon lit: Battery is low. Lo icon lights up: battery is exhausted. Please connect adapter to charge. ※ Do not use this product while charging.
a.Insert the air tube plug into the air tube connector a. Make sure the arm strap connector is fully inserted into the Upper Arm Blood Pressure Monitor vent. b. Please take off your coat, sweater and other thicker clothes, do not have any accessories, bare upper arm or wear a thin shirt for measurement. c. Wrap the cuff around the left arm. d. Do not wrap the cuff too tightly (It is the best to easily insert a finger) and the lower edge of the cuff should be 2~3 cm from the hand bend (as shown). e.
measurements Note: Comparative blood pressure measurements should be performed in the same condition (usually referred to quiet conditions)! If the arm artery is low (high) relative to the heart , it will result in a high (low) blood pressure value. (Each 15cm height difference will produce 1.3kPa/mmHg error value)! This unit is not suitable for newborns. Note: Please use the original arm cuff that meets the clinical test requirement! The loose or air bag exposed cuff will lead to false blood pressure.
Stop measurement If there is any discomfort or for some reason, such as when the airbag is in a persistent over-inflated state, there may be a risk. At this time, the blood pressure measurement must be stopped. You can press the [START/STOP] button and the Upper Arm Blood Pressure Monitor will immediately release the air pressure in the cuff to suit your need. 2 Confirm measurement results Automatic storage of measurement results. SYS 1) irregular DIA pulse: PULSE The unit can detect the arrhythmia.
Press the [MEM] button to display the latest measurement results Query memory value 9.2 Delete memory This Upper Arm Blood Pressure Monitor can only delete all records at one time. Cannot delete a single record. When the measurement result is displayed, press and hold the [MEM] key while pressing the [start/stop] key for about 3 seconds. ※Please do not press the 【start/stop】key first, otherwise the power will be cut off. 10.
1. In shutdown mode, press the [SET] button to switch to user 2, which can be switched cyclically. 10.2Setting time, date and kPa/mmHg unit selection The unit automatically records the date and time of each measurement, which is very important.Because the blood pressure of the human body is constantly changing on the same day, it is recommended that you set the correct date and time immediately after installing the battery.
4Press the [SET] button again, the screen will switch to the time setting. At this time, the first digit (indicating hour) starts to flash. Please input the time by pressing the [MEM] button (for example, you can change the time from 0:00 to 12:00 by pressing the [MEM] button 12 times). 5.Press the [SET] button again, and the last two digits (in minutes) will start to flash.
6.Press the [SET] button again to enter the kPa/mmHg unit selection. You can select kPa or mmHg by pressing the [MEM] button, “0.0” will display on the screen when kPa is selected, and “0”will display when mmHg is selected. 8. In the last step, press the [SET] button, and the date will be displayed after the setting is completed (the date will be displayed first, and it will automatically jump to the time after 3 seconds). When there is no operation, it will automatically shut down after 30 seconds. 11.
The meaning of the indicator color: Green, it represents normal blood pressure. Red,please consult physician 12.Maintenance and storage 11 Remove the plug of the air pipe 2 Gently bend the air pipe of the arm band and put it into the arm band Do not over bend the windpipe of the arm band. Otherwise it may not be possible to make a correct measurement.
disinfection. If it is heavily soiled, it is recommended to replace the cuff. You can contact the dealer or manufacturer for disposal. This product is intended for home use and is a personal product. Please use medical alcohol to clean it before use it to others. Note: This product is not waterproof. Please be careful not to be too moist when cleaning. Also it is not suitable for splashing with water. Please do not use any other solutions to clean this unit except those method recommended by manufacturer.
Q: Why is the blood pressure value measured at home lower than the blood pressure value measured in the hospital? A: I often feel nervous when measuring blood pressure in the hospital, so the measured value will be high. Because I feel calm when measuring in my own home, sometimes it is 20mmHg~30mmHg (2.7kPa~4.0kPa) lower than the hospital's measurement value. It is very important to know the blood pressure measurement value when you are calm at home.
Q: What is the key to blood pressure management? A: Please record your medication status and living conditions.Recording daily changes in blood pressure can provide a clearer picture of blood pressure trends, which can be more helpful in personal health management.In addition, this record is very important for the doctor's diagnosis.
At the same time, the United Nations Committee on the Investigation, Evaluation and Treatment of Hypertension in 1988 recommended that whether it is systolic (SYS) or diastolic (DIA), average person must undergo at least three measurements ,further diagnosis is needed when blood pressure is found to be higher than the normal level. 14.2 What measure should be taken when blood pressure is too high or low? a) Please consult physician.
factory Er01 Error of Too much noise to detect an Please rest for 2 minutes, adjust the cuff, keep measurement effective pulse signal quiet during measurement, do not move or No pulse was detected talk. Error of measurement Error measurement of Test results deviate from normal range Static pressure Pressure exceeds design ceiling exceeds the set protection Measure blood pressure again protection point Zero overtime 1. During the process of returning to 1.
If any faults (or abnormal conditions) occur during use, you can check and exclude according to the items listed in the following table: Malfunction When the battery is installed and the switch is turned on, Exclude Check whether the product has power shows nothing. The air pump has begun to inflate, but there is no rise in * Check the hose connection for air leakage or whether it is fully arm pressure. inserted into the socket. Upper Arm Blood Pressure Monitor s fail to measure blood 1.
Mode FDBP-A10BL display LCD display Measuring method Oscillometric method Memory Automatically store each 60 sets of measured values for two users Resolution 0.1kPa(1mmHg) Measurement Range stress: 0kPa~39.3kPa(0~295mmHg); Pulse: 40 to 199beats/min Accuracy Static pressure: ±0.4 kPa (±3 mmHg) Pulse: within ±5% of the reading Power 3.
20.
21.Electromagnetic compatibility information FCC STATEMENT This device complies with part 15 of the FCC Rules. Operation is subject to thefollowing two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation FCC CAUTION Changes or modifications not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment.
Warning: ·Use of this equipment adjacent to or stacked with other equipment should be avoided because it could result in improper operation. If such use is necessary, this equipment and the other equipment should be observed to verify that they are operating normally.
Floors should be wood, concrete or ceramic Electrostatic ± 8 kV contact ± 8 kV contact discharge (ESD) ± 2 kV, ± 4 kV, ± ± 2 kV, ± 4 kV, ± IEC 61000-4-2 8 kV,± 15 kV air 8 kV,± 15 kV air tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. If ESD interfere with the operation of equipment, counter measurements such as wrist strap, grounding shall be considered.
Conducted RF 3 Vrms 3V IEC 61000-4-6 150 kHz to 80 MHz 150 kHz to 80 MHz 6 V in ISM and 6 V in ISM and amateur radio amateur radio bands between 0,15 bands between 0,15 MHz and 80 MHz.a 80 MHz to 800 MHz MHz and 80 MHz.a Radiated RF 10 V/m 10 V/m IEC 61000-4-3 80 MHz to 2.7 GHz 80 MHz to 2.
a The ISM (industrial, scientific and medical) bands between 150 kHz and 80 MHz are 6,765 MHz to 6,795 MHz; 13,553 MHz to 13,567 MHz; 26,957 MHz to 27,283 MHz; and 40,66 MHz to 40,70 MHz. The amateur radio bands between 0,15 MHz and 80 MHz are 1,8 MHz to 2,0 MHz, 3,5 MHz to 4,0 MHz, 5,3 MHz to 5,4 MHz, 7 MHz to 7,3 MHz, 10,1 MHz to 10,15 MHz, 14 MHz to 14,2 MHz, 18,07 MHz to 18,17 MHz, 21,0 MHz to 21,4 MHz, 24,89 MHz to 24,99 MHz, 28,0 MHz to 29,7 MHz and 50,0 MHz to 54,0 MHz.
Famidoc Technology Co., Ltd. Add.: No. 212 Yilong Road, Hexi Industrial Zone,Jinxia, Changan Town, Dongguan 523853, Guangdong Province, P.R. China. Tel.: +86-769-89272488 Website: www.famidoc.com Name: Shanghai International Holding Corp. GmbH (Europe) Add: Eiffestrasse 80, 20537 Hamburg, Germany PN: Description Version: V1.