Co Camera Accessories User Manual
SECTION 3 
Camera Control 
3.1 Quick Start with the CameraLink
TM
 Interface 
Follow your Frame Grabber manufacturer’s instructions when inserting the frame 
grabber card into your PC. Be sure that you install the correct configuration 
software that matches the camera you are using. 
At this point your camera should already be connected to the DC power supply and 
connected to the frame grabber in your PC. 
To communicate with your camera we have created specific commands. In the next 
section we will go over these commands in detail. 
3.2 Camera Control Signals using Camera Link™ Interface 
Four LVDS pairs are reserved for general-purpose camera control. They are defined as 
camera inputs and frame grabber inputs. In the case of Fairchild Imaging’s CAM/CCD-
2KCL.TDI and CAM/CCD-4KCL.TDI cameras the signals are: 
Line Sync (generically Camera Control 1, CC1) 
Frame Sync (generically Camera Control 2, CC2) 
Spare, unused (generically Camera Control 3, CC3) 
Spare, unused (generically Camera Control 4, CC4) 
3.3 Serial Communication 
The serial interface for the camera has the following characteristics: default baud rate 
38400, one start bit, one stop bit, no parity, and no handshaking. 
3.4 Commands 
Note: Commands are not case sensitive. To input commands, type in the three 
character command and hit enter. 
 Command Syntax: 
Command + enter, example: Type in RSH, then enter 
Command + enter + value, example: Type in SSM then enter, then type 2 
Command + enter + value + enter, example: Type in SLR then enter, then type 
2000, then enter 
Fairchild Imaging • CAM/CCD 2KCL.TDI & CAM/CCD-4KCL.TDI Line Scan Camera User’s Manual • Rev C• 16 of 30 