Section 6: Service & Maintenance
TROUBLESHOOTING
Important Precaution
Before handling the FB3000 boards or other system
components, a service technician must always be properly
grounded.
Electrostatic Discharge (ESD) can easily damage the
FB3000 board assemblies.
Remove the anti-static packing only when the parts are
ready for installation.
Handle the boards by their edges, and avoid touching their
components.
Error Condition(s) Solution(s)
Check that scale is empty.
If Scale is empty, Call for
Service.
Load Cell(s) bad.
A large amount of weight is zeroed.
This is normal.
Press
OK and continue weighing.
Possible load cell damage.
Call for Service.
Load Cell Failure(s)
Flashing and displays “
- - - -
Possible load cell damage.
Call for Service.
Access the
Load Cell Diagnostics Menu to verify the
load cell status.
Count stability or change of counts.
Contact the local service for further trouble-shooting.
SC Cells Found None
Possible damaged load cell cable.
Load cell shortened.
Defective Pit Power Supply.
Defective Smart Sectional Controller(s).
Defective Analog Assembly.
Displays “ - - - - “ ~ lb GROSS
Communication error to load Cells.
Check settings by pressing
F10.
Settings should be COM2, Even.