3.5.1. I/O MENU DESCRIPTIONS, CONTINUED
PROMPT COMMENT(S)
Port A
Port A has available selections of the following:
UnUSEd - no outputs
Printr - printer output
Contin - continuous output
noALog, AnALog
Used to turn on or off the 4 - 20MA output.
noALog = Off
AnALog = On
Protob
Protocol setting for Port A choices include the following:
bAUd 9600, 600, 1200, 2400, and 4800
ChAr 8 and 7
PAritY nonE, odd, and EuEn
tArgEt
Conditional, if setpoints are selected in Port B.
1. Press
ENTER to adjust the Setpoint Target Weight Menu.
─ The large display shows the Current Target Weight Value
associated with the Target Number, as shown in the small
display.
─ The
SCROLL key toggles through available Setpoint Relays
(Targets).
2. Enter a new Target Weight Value with the numeric keypad.
3. Press
ENTER to accept the new weight value and advance to the
next available target.
Port b
Port b can be configured for the following:
UnUSEd - no outputs
* SEtPnt - Setpoint modes
* PoLL or PoLL id - computer output
Printr – printer
Contin - continuous output
Poll
Used for Demand Output to a computer.
When the instrument receives an upper-case "W" followed by a
carriage return ("W", "Cr") from the PC, a data stream transmits.
See Appendix VII for ASCII Chart.
Poll id
Use this demand mode when a specific instrument ID is required.
Setup by entering the required ID's decimal equivalent.
See Appendix VII for ASCII Chart.